Setup
1.
Remove the instrument from the crate and place it on a flat, stable
surface.
2. Turn off the Nitrogen valve (horizontal). Turn off both Main Power
switches and all four Heat switches. Make sure all regulators are backed
off completely. Turn them completely counterclockwise.
3. Connect a Nitrogen source to the back of the unit. The inlet connector is
¼" NPT.
4.
On the back of the unit is a power inlet configured for 115 VAC or 230
VAC. Plug the supplied power cord into an appropriate power source.
- #170-00-4S:
115 VAC, 50/60 Hz
- #170-00-4S-230: 230 VAC, 50/60 Hz