HTHP Filter Press for Cement Testing

 

with 175-mL, Double-Capped Test Cell and N

2

 Pressuring Manifold

#170-00-2: (115 V) 

#170-01-2: (230 V)

Instruction Manual 

Updated 7/31/2017 

Ver. 3.0

OFI Testing Equipment, Inc. 

11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A. 

Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com

 

©

Copyright OFITE 2015

Summary of Contents for 170-00-2

Page 1: ...pped Test Cell and N2 Pressuring Manifold 170 00 2 115 V 170 01 2 230 V Instruction Manual Updated 7 31 2017 Ver 3 0 OFI Testing Equipment Inc 11302 Steeplecrest Dr Houston Texas 77065 U S A Tele 832 320 7300 Fax 713 880 9886 www ofite com Copyright OFITE 2015 ...

Page 2: ...1302 Steeplecrest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 1 Intro 2 Specifications 2 Components 3 Safety 5 Diagrams 6 Operation 7 Warranty and Return Policy 11 Table of Contents ...

Page 3: ...ng A controlled Nitrogen pressure source not included A pressure manifold with two regulators primary and back pressure A high pressure test cell with removable cement screens A heating jacket for heating the test cell A suitable stand Size 7 5 11 23 5 19 1 27 9 59 7 cm Weight 27 lb 12 3 kg Shipping Size 17 23 5 12 5 43 2 59 7 31 8 cm Shipping Weight 38 lb 17 2 kg Maximum Temperature 400 F 204 C M...

Page 4: ...wer Cord 115V Only 170 06 1 Back Pressure Receiver 15 mL Stainless Steel Tube for N2 144 11 90 Street Ell 144 15 Plate Brass Bushing NPT Male to NPT Female 170 07 O ring 170 28 Receiver Body 170 32 Needle Valve Male NPT 171 23 1 Safety Pin with Lanyard 170 19 Filter Paper 2 6 35 cm Specially Hardened for Filter Presses 170 35 6 Adjustable Wrench 170 45 Test Cell Assembly Double Capped for Cement T...

Page 5: ...PSI 2 NPT Bottom 171 90 06 Reducing Bushing MNPT FNPT 316 Stainless 171 90 07 Hex Nipple NPT Qty 2 171 90 13 Fitting Flare Male NPT Optional 170 37 Nitrogen Cylinder 21 7 Right hand Thread 170 00 2 SP Spare Parts for One Year 143 00 1 Diaphragm for Airco Regulator 143 07 Repair Kit for 143 00 Regulator 153 14 Glass Graduated Cylinder 50 mL 1 mL Qty 2 154 10 Dual Scale Thermometer with Metal Dial 5...

Page 6: ...rature high pressure HTHP filtration Under high temperature and pressure Nitrous Oxide can detonate in the presence of grease oil or carbonaceous materials Nitrous Oxide cartridges are to be used only for Garrett Gas Train Carbonate Analysis Due to the high temperatures and pressures involved in this test extreme care must be exercised at all times All safety precautions must be met especially in ...

Page 7: ...ver 170 06 1 Bleeder Valve 170 32 O ring 170 13 3 Locking Screw 170 26 1 Valve Stem 170 16 Screen 170 18 Safety Pin 171 23 1 With Lanyard Bleeder Valve 170 32 Manifold Block 170 20 Safety Pin 171 23 1 With Lanyard Outlet Valve Stem 170 16 Nitrogen Manifold Primary Pressure Gauge 171 40 Source Pressure Gauge 171 42 Back Pressure Gauge 171 38 Connect to Back Pressure Receiver Manifold Block 170 20 ...

Page 8: ...ll making sure the arrow on the cap lines up with the arrow on the cell body Tighten the locking screws with the supplied allen wrench Screw a valve stem into the cell cap and tighten it completely Standard Buna N o rings 170 13 should be used only for tests up to 300 F 149 C For tests over 300 F 149 C use Viton o rings 170 13 3 After tests that come close to the maximum temperature it will probab...

Page 9: ...ting as the valve stems are opened and closed Transfer the thermometer from the heating jacket to the thermometer well within the cell 6 Connect the pressuring assembly to the top valve stem and lock it in place with the retaining pin Place the back pressure receiver on the bottom valve assembly and also lock it in place with the retaining pin Cap Locking Screws 170 26 1 Arrows 7 Unscrew the T scr...

Page 10: ...essure Minimum Back Pressure F C PSI kPa PSI kPa 212 100 14 7 101 100 690 250 121 30 207 100 690 300 149 67 462 100 690 350 177 135 932 160 1 104 400 204 247 1 704 275 1 898 500 260 680 4 692 700 4 830 For tests above 400 F use Teflon o rings 9 When the fluid sample reaches the desired test temperature increase the primary pressure the regulator on the left to 1 000 PSI 3 448 kPa more than the bac...

Page 11: ...0 minutes in cool water before loosening the cap locking screws and removing the cell cap 13 Correct the total filtrate volume collected to a standard filtration test area of 7 1 in2 45 8 cm2 by doubling the filtrate volume collected in 30 minutes Record this total filtrate volume doubled temperature pressure and time 14 Place the cooled cell upright with the outlet cap side or filter side down Op...

Page 12: ...any losses or damages that occur as a result of Improper installation or maintenance of the products Misuse Neglect Adjustment by non authorized sources Improper environment Excessive or inadequate heating or air conditioning or electrical power failures surges or other irregularities Equipment products or material not manufactured by OFITE Firmware or hardware that have been modified or altered b...