Maintenance
Slurry Cup
After every test, immediately disassemble the slurry cup and clean it
thoroughly with soap and water. Be sure to remove any residual cement
before it hardens. Hardened cement on any of the parts can cause
irreparable damage.
Test Cell
After every test, examine the inside of the test cell for any cement or other
debris. If necessary, wipe the inside of the cell with a rag or paper towel.
Oil Filter
If oil is not flowing from the oil reservoir to the pump, it may be necessary
to replace the oil filter.
1. Open the back panel of the main unit cabinet and locate the oil
filter fixture.
2.
Open the fixture, remove the filter, and replace it with a new one.
To reduce the risk of damaging the piping around the fixture, you
can completely remove the entire fixture from the unit. Once you
have replaced the filter, be sure to re-install the fixture with the oil
flow directed downward. An arrow printed on the side of the fixture
shows the direction of the oil flow.
Oil Filter Fixture
Fixture
Filter