3. To engage the drive bar of the slurry cup into the potentiometer, rotate the
cup with the motor for a few seconds while applying slight pressure to the
potentiometer. Note that if the unit is in an alarm condition the motor will
not engage.
4. Place the metal seal ring, the o-ring, and the retaining ring onto the cell
cap as shown below. Place the cell cap onto the cell and hand tighten.
Thermocouple
Cell Cap
Thread Gland
5. Plug the thermocouple into the port on the unit cabinet. Insert the
thermocouple into the hole in the top of the cell cap and tighten the thread
gland finger tight. Then loosen it ⅛ of a turn.
6.
With a ⅝" wrench handy, turn the “Oil Reservoir Valve” to “Fill Cell”. The
test cell will begin to fill with mineral oil from the reservoir. Carefully watch
the top of the test cell. When oil begins leaking out of the thermocouple
hole, tighten the thread gland with the wrench. This will ensure that no air
remains within the test cell.
Retaining Ring
(#120-148)
Metal Seal
(#120-90-049)
O-ring, Viton
(#120-149)