U S E R   M A N U A L   D A Y C O R ® S U P E R B  

 8 

 

How to count 
 

Enter the counting mode by pressing on the counting function-key. A small frame will 

appear on the display. By pressing the up and down arrows the size of the frame will 

change. Position the corona image in the center of the frame and wait until the count 

value stabilizes with only small fluctuations. 

 

To count accurately

 

1)

 

Set the UV gain default to the lower value recommended by the manufacturer, 

as appears in your supplied instruction manual.  

2)

 

Verify that the camera operates at normal video rate – LI0

 

(no long integration)

 

in the status line. 

3)

 

Set the camera to manual focus and defocus. By defocusing  the camera will 

count UV spots that were overlapping in the focused mode and were counted 

as a single 

spot. 

4)

 

The displayed 

counting is an 

average of the 

last 4 display 

values, it is 

therefore 

necessary to wait 4 successive displays, about 25 sec. 

Compare counting

 

Counting can be used as a comparative tool. When comparing counting, it is 

important to compare just the reading but keep all other variables that affect 

counting such as distance, angle, same camera, and same ambient conditions – 

equal.  

 

(*)

 The camera is not sold as an electrical measuring device and its output does not read directly the pC value of a corona 

 

 

Exposure  

 

The visible camera, by default, is set to automatic exposure mode.  

Under certain light conditions details like cables may be invisible and changing 

manually the exposure might help. The exposure affects the recorded video clips and is 

a function of the visible channel only.  

Focused 

Defocused 

Summary of Contents for daycor superb

Page 1: ...1 DayCor Revision June 2011 Ofil Ltd User s Guide...

Page 2: ...POB 4016 Nes Ziona 74140 Israel 16 Einstein Street Weizmann Science Park Nes Ziona Israel 74000 www ofilsystems com ofil ofilsystems com Tel 972 8 940 7953 Fax 972 8 940 7873 USA Toll free 1 866 27986...

Page 3: ...G 6 UV GAIN 6 LI LONG INTEGRATION 7 COUNTING 7 EXPOSURE 8 STATUS DISPLAY 9 LCD BACK LIGHT 9 ZOOM 9 VIDEO RECORDING 10 STILLS RECORDING 10 FIRST TRYOUT 10 CHAPTER 3 11 INSPECTION TIPS 11 HANDLING NON C...

Page 4: ...just the shoulder straps to fit your body by fastening the straps 3 Fasten the front pad 4 Lean the camera against the front pad Use the grooves to adjust the viewing angle of the camera 5 hook the su...

Page 5: ...ation and for report generation Here is an example of a status line with and without counting Option 1 with counting COMBINE AF GAIN 180 COUNTER 10000 Option 2 without counting COMBINE AF GAIN 180 LI...

Page 6: ...e focus scale is also used when the camera needs to be recalibrated a procedure that can easily be performed by the camera operator special instructions When to use manual focus When Corona is out of...

Page 7: ...m UV signals can be removed while the repetitive signal although decreasing in size will still be visible on the screen More in Chapter 3 Counting The counting function provides an indication of the c...

Page 8: ...spots that were overlapping in the focused mode and were counted as a single spot 4 The displayed counting is an average of the last 4 display values it is therefore necessary to wait 4 successive di...

Page 9: ...or keeping logs follow up documentation etc To turn off press on the OSD key LCD Back Light The LCD back light can be adjusted by the dedicated knob on the LCD Note that the brightness does not affect...

Page 10: ...folders that are created automatically by date and time Use arrows to navigate withing a folder Display an image using the F1 Step back using F2 Leave playback by pressing any other button First tryo...

Page 11: ...r the feeling as if the camera is not solar blind Although those random UV spots do not harm the camera it is recommended not to look for a long time directly on a welding machine with the camera A st...

Page 12: ...the UV channel of the DayCor Superb camera turns into a slow response instrument If the camera moves relative to the UV source the UV image will be smeared When recording while operating in the LI mo...

Page 13: ...center or off center Or else consult the manufacturer Focus If the camera needs to be recalibrated ask the manufacturer for the procedure to be performed by the camera operator To check alignment of t...

Reviews: