-TRANSMITTER 4/20mA series TS-TE-T4-TX 
 

       x TS-T4                                                                   x TE-TX 

 
                                                                                                                                                                      

wiring 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

5 SETTING AT WORK 

Make sure that the use of the instrument does not exceed the intended use (i.e. pressure and temperature values and the 
like) and ensure the measurement made is correct by changing a few times the level of the fluid in the vessel. 
If this cannot be done, you can move the float by hand. 
As  far  as  the 

version  with  transmitter  is  concerned

,  series  connect  an  ammeter  in  the  line  to  check  it  for  proper 

operation. 
The standard setting ranges from 4mA = 0% to 20mA = 100% (on request, the setting can be reversed). 
Comply with the power supply value 10-35 Volt for standard models and d10-28 Volt for explosion-proof models. 
The maximum load is given by the following formula: 

 
 

R

 LOAD

  

= (V 

SUPPLY

 – 8) / 0.023 

 
The instrument shall never work close to strong magnetic fields (minimum distance: 1 m) 
It can only be used with special feeders or indicators. 

 

As far as the 

version with alarms 

is concerned, check the alarms for proper operation by making the float slowly slide 

along the rod to make sure the alarm switching takes place. 
 

6.CALIBRATION 
Adjusting the pointing head 

Make  sure  the  float  is in the  “minimum level”  position, then remove  the  front lid and loosen the  hand fastening screw 
located  beneath  the  latter.  Align  the  hand  on  the  “ZERO”  or  minimum  level  position  if  other  than  zero);  tighten  the 
locking pins taking care not to damage the hand. 
IN  the  models  equipped  with  two  hands,  (RED  HAND  =  metres  and  BLACK  hand  =  centimetres),  in  tightening  the 
locking pins, take care to maintain the hands on the fastening pin correctly spaced from one another, to prevent them from 
locking out while turning. 

 
 
 
 
 
 
 

Summary of Contents for TLQ Series

Page 1: ...not require any power supply 2 MODEL IDENTIFICATION TLQ TRANSMITTER Hart Transmitter Hart Exia Transmitter Exia IIC T6 transmitter No trasmitter Std Transmitter TX T0 TS TE GAS thread NPT thread UNI...

Page 2: ...t 4 to move by magnetic coupling The movement of the rope which is wound around a pulley 6 with a return device causes the hands of the clockworks assembly to move and indicate the fluid s level in th...

Page 3: ...een the magnets 4 Release the wire very slowly It will wind round the pulley until it gets tightened 5 Screw the instrument s neck onto the coupling flange 6 Remove the round cover opposite to the poi...

Page 4: ...made through cable entrances or tube piping in compliance with the EN 60079 14 standard The cable entrance shall not affect the specific properties of the protection method as indicated in the EN 600...

Page 5: ...y value 10 35 Volt for standard models and d10 28 Volt for explosion proof models The maximum load is given by the following formula R LOAD V SUPPLY 8 0 023 The instrument shall never work close to st...

Page 6: ...INTENANCE Mechanical dial type level indicators do not generally require any scheduled maintenance Periodically check once every six months approximately that the rod and the float are clean and the i...

Page 7: ...DRAWING Dimensioning values required C Connections L Range E Extension F Float D Extension connection 9 SPARE PARTS A THREADED THREADED F THREADED THREADED FLANGED FLANGED TLQ H TLQ S 160 220 UNI DN 2...

Page 8: ...ons in force All metal parts after removal of seals and gaskets special protective coatings requested by the customer and all other plastic parts can be recycled 12 GUARANTEE All level indicators are...

Page 9: ...clean and free from dust or deposits Otherwise OFFICINE OROBICHE reserves the right not to carry out the servicing and return the instrument to the sender FINAL REMARKS Each instrument is supplied ful...

Reviews: