Computer interface
GA 09.035/6.02 - 07/2011 - © Oerlikon Leybold Vacuum
73
6.3.19
OFD
Offset display. Offset correction values for linear transmitters (CTR). See
Chapter 5.2.4 Offset (oFS),
S: OFD[,±a.aaaaE±aa,±b.bbbbE±bb,±c.ccccE±cc] <CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: ±a.aaaaE±aa,±b.bbbbE±bb,±c.ccccE±cc<CR><LF>
6.3.20
PNR
Program number. Firmware version number. See Chapter 5.4.2 Firmware ver-
sion (Pnr),
52.
S: PNR<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a<CR><LF>
Parameters
Significance
±a.aaaaE±aa
Offset value of transmitter 1 in current unit of
measurement
Default: 0.0000E+00
±b.bbbbE±bb
Offset value of transmitter 2 in current unit of
measurement (see above)
±c.ccccE±cc
Offset value of transmitter 3 in current unit of
measurement (see above)
Parameters
Significance
a
Firmware version
Example: 302-533-F