OEC UroView
®
2800 Table/Generator Functional Block Diagrams
55
Periodic Maintenance
Page 2 of 2
1A
Interconnect
Plug
Back
Plane
1
KVp and mA changes are sent to the generatorvia the CAN Bus.
Camera Gain control is sent through the Manchester communications to the CCD camera..
See KV and MA Functional Diagrams.
2
Table/Generator
Interface PCB
CCD
Camera
Image
Intensifier
XRAY Tube
Generator
2
P3
C
Cam_Comm-
P8
ISA
Arcnet
Node
11
P1
P3
2
1
Arcnet_HI
Arcnet_LO
8
7
ABS Path
Control Path
07/01
2800
GE OEC Training
g
28ABS.DS4
U16
U14
U56
U22
CAN
Interface
U24
U19
U36
Decoder
J1
30
12
3
4
6
5
ISA
Half Size
CPU
J6
C
Cam_Comm-
CAN_GEN_H
CAN_GEN_L
Generic
Interface
Board
P2
33
15
11
ABS Path