Page 2 of 2
For Reference Only
3/10/97
IRIS_C.DS4
COMPACT 7600
IRIS MOTOR DRIVE LOGIC
IRIS COLLIMATOR CONTROL
NORMAL
MAG
HAND
R29
1
7
7
1
7
7
1
1
7
7
1
NC
NC
14
7 14
NC
NO
NO
R30
R58
NC
NC
NO
7 14
The desired Iris
position signal
out of K2-7 is
sent to the Iris
comparator circuit
+
-
+
-
+12V
U6
U7
2
3
5
5
2
3
IRIS CLOSE
IRIS OPEN
R34
R52
K7
9
11
13
8
6
4
-MIBZ
-MIBA
K6
K5
6
2
7
14
14
2
7
6
B101 BOARD
D
esi
re
d I
ris
p
os
itio
n
13
U1
IRIS
CONTROL
LOGIC
PAL U2
PAL U3
TRANS T7
NAND U9
R36
Manual
Iris speed
adjustment
+RBLMIN
+RBLPLU
U1
4
3
14
C20
A20
A21
R2
POTIM
+12V
LIMIT SWITCH
B45
BOARD
FILTER
BOARD
B1OO
BOARD
XP3
P2
X3
P3
DRIVE
CIRCUITS
10
5
7
10
11
12
10
10
11
11
12
12
10
11
12
A21
A20
C20
ST1
ST4
XRAY HEAD ASSEMBLY
IRIS CLOSE
IRIS OPEN
MP3
Relay contacts (K1, K2, K8) are shown
in the proper condition for each respective
function unless indicated otherwise
K5,K6 & K7 are shown in
the de-energized condition
LIMIT SWITCH
16V
-MIBA
-MIBZ
OPEN
CLOSE
K5
K6
1
2
Aprox 12.5VDC (chassis Gnd) measured with a DVM when active (close)
16V
0V
0V
measured with a scope when active (close)
Aprox 12.5VDC (chassis Gnd) measured with a DVM when active (open)
16V
0V
0V
measured with a scope when active (open)
IRIS
MOTOR
ST1
(23 cm / 9 inch)
(15 cm / 6 inch)
(15 cm / 6 inch)
Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800
An Independent C-Arm Service Provider. This PDF provided for research / historical purposes only.