FLUORO MODE CONTROL LOGIC
3/7/97
Fluoro_C.ds4
CONTROL PANEL
U2
FLUORO EXPOSURE TIME
DISPLAY
ST3
Xray Indicator
H3
H4
H5
H6
10
Min
1
Min
10
Sec
1
Sec
6
7
8
9
10
11
12
13
16V
-XRAY
16
OPTO
ISOLATOR
U8
HAND SWITCH
FOOT SWITCH
-HAND
-FUSS
OPTO
ISOLATOR
U9
3
6
6
3
-HANDSW
-FOOTSW
ST20
ST10
B106 BOARD
2
1
2
1
D15
CONTROL
LOGIC
PAL U2
PAL U3
U11, U12
RELAY K1
2 second
TIMER U14
OPTO
ISOLATOR
U3
OPTO
ISOLATOR
U3
OPTO
ISOLATOR
U2
OPTO
ISOLATOR
U7
-HANDSW
-FOOTSW D14
ST6
5
2
-FLUOREQ*
-MEMFLUREQ
D53
MONOTOR
XRAY LAMP
LOGIC
RELAY K1
INVERTER U8
CONTROL
LOGIC
PAL U1
B100 BOARD
B104 BOARD REV A
-FLUOREQ*
-MEMFLUREQ
-XRY
-STRA
-SPT
-SPR
-TRIGIN
-TRIGOUT
2
8
16
2
26
6
21
44
4
16
14
2
U11-10
U3-4
ST5
ST1
11
10
11
10
U8-8
24VAC
K1-1
From B111MR board X4-2
(see Monitor Rotation diagram)
To/from image processor
MONITOR ASSEMBLY
XRAY TUBE HEAD
ASSEMBLY
FI
LT
ER
BO
AR
D
XRAY LAMP
B103 BOARD
ST4
4
4
X1
X4
1
24VAC
X3
3
4
10ER
TIMB
100ER
1000ER
1ER
TIMA
TIMD
TIMC
6
7
8
9
10
11
12
13
ST2
B101 BOARD
FLUORO TIME
DISPLAY LOGIC
PAL U17
PAL U18
U4, U10
U11, U12
U13, U14
Jumper BR2
IN = 4 sec fluoro beep
OUT = no flouor beep
Jumper BR1
Always out
Jumper SCH1
1 2 3
-XRAY
B64 BOARD
B143 BOARD
K1-14
ST1
16
18C
-X
R
A
Y
-X
R
A
Y
X1
18C
X1
11A
ST1
18C
ST4
-XRAY
-XRAY
3
4
TRIGIN
= enables the image processor.
TRIGOUT
= from the image processor stays active, enabling exposure, for
640 msec after the xray switch is released. This ensures sufficient time to for last
image hold. The timer U14 on the B100 board terminates exposure after 2 seconds
if the TRIGOUT does not terminate after the 640 msec time period.
1
ST1
17
16V
LED 4
S7
Auto
Fluoro
Mode
-KFLUORO
U12-4
U12-2
-KFLUORO
17
ST2
D20 U12-15
U23-1
U10-6
16VDC
D43
Xray-On (Green)
5 Minute Warning Indicator
16V
-5MINLED
36
S28
35
-5MINSW
36
35
29A
29C
U10-4
U18-23
ST1
29C
29A
-LED
ALAR
M
-C
LR
B
U
Z
-GERR
(Generator Error)
(See page 2 for REV B board)
(Using the B104 Rev A board)
.
F1
1A
Power for board
ST1
23C
-BUZZER
Mother
Board
ST9
U11-2
COMPACT 7600
SEE PAGE 2 FOR SYSTEMS WITH B104 Rev B
Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800