background image

2 mm

L

www.odu.de

ODU MEDI-SNAP® 

Page 10

Insulation

Cable

Notes

3. Recommended adhesive for the back nut

1. Recommended stripping length

Scotchweld

 (grey)

, ODU part number 890.204.000.030.025 

Recommended cleaning agent: Isopropyl alcohol.

Caution! If adhesives that have not been released are 

used, cracks may appear after some time. Use only the 

indicated adhesive.

2. Tightening torque

Dimension L

mm

Size 1

Size 2

12

15

Torque moment

Size 1

Torque moment

Size 2

Nm

Nm

Straight plug

0.25

0.5

Right-angled plug clamping cap

0.30

Right-angled plug clamping nut

0.25

In-line receptacle

0.25

4. Tools/Accessories

 

– ODU open-ended spanner see ODU MEDI-SNAP  

product catalogue section accessories and tools

 

– ODU crimping tool see ODU MEDI-SNAP product  

catalogue section accessories and tools.

010.013.005.000.000 

Assembly Instructions

Summary of Contents for MEDI-SNAP

Page 1: ...www odu de ODU MEDI SNAP Assembly Instructions Straight Plug Size 1 Right angled Plug Size 1 In line Receptacle Size 1 Straight Plug Size 2...

Page 2: ...nd cable collet over the cable 2 Strip cable and wire see page 10 3 Pre tinning of strands 4 Solder the litz wires A according to contact arrangement Recommended cleaning agent Isopropyl Alcohol 010 0...

Page 3: ...ntacts 6 Put the assembled cable con sidering the guidings B into the housing 7 Screw and fasten back nut on the assembled in line receptacle Caution Consider tightening torque see page 10 If necessar...

Page 4: ...wires according to contact arrangement Recommended cleaning agent Isopropyl Alcohol 5 Slide the right angled part against contact insert Caution Don t damage the contacts 6 Slide the cable collet aga...

Page 5: ...angled part and screw it with the back nut If necessary secure thread C with adhesive see page 10 Caution Consider tightening torque see page 10 8 Screw back screw on the assembled plug and fasten it...

Page 6: ...ire see page 10 3 Pre tinning of strands 4 Solder the litz wires A according to contact arrangement Recommended cleaning agent Isopropyl Alcohol 010 013 004 000 000 Assembly Instructions Housing Cable...

Page 7: ...ntacts 6 Put the assembled cable con sidering the guidings B into the housing 7 Screw and fasten back nut on the assembled in line receptacle Caution Consider tightening torque see page 10 If necessar...

Page 8: ...cable and wire see page 10 3 Pre tinning of strands 4 Solder the litz wires A according to contact arrangement Recommended cleaning agent Isopropyl Alcohol 010 013 005 000 000 Assembly Instructions Ho...

Page 9: ...ontacts 6 Put the assembled cable considering the guidings B into the housing 7 Screw and fasten back nut on the assembled in line receptacle Caution Consider tightening torque see page 10 If necessar...

Page 10: ...after some time Use only the indicated adhesive 2 Tightening torque Dimension L mm Size 1 Size 2 12 15 Torque moment Size 1 Torque moment Size 2 Nm Nm Straight plug 0 25 0 5 Right angled plug clampin...