background image

Page 

10

 of 

13

 

Section 3 - Specifications for DLS350 & DLS355 

 

 
 

 

 
 
 
 
 
 
 
 
 
 
 

 

 

  

    

 

Specifications

 

DLS 350 

DLS 355 

Wavelength 

850nm & 1300nm LED 

1310nm & 1550nm Laser 

Output power 

- 20dBm Typ. 

- 4.0 dBm Typ. 

Laser Classification 

NA 

Class 1 (FDA 21 CFR 1040, 11) 

Output stability 

+/- 0.05dB, one hour 
+/- .03dB long term after 15 minute warm-up 

Spectral Width 

40nm/120nm 

5nm/5nm 

Optical interface 

Universal connector 
(contact ODM for other connector styles) 

Tone Output 

2kHz 

Power on 

Push button ON, Auto shutoff 

Battery 

CR 2 , Optional AC power for long term 

Dimensions 

6.1x.94x.75 in. (15.5x2.38x1.90cm) 

Weight 

3.0 oz (8.5g) 

Operating Temperature 

-10 to 50˚C 

Storage Temperature 

-30 to 60˚C 

Summary of Contents for DLS350

Page 1: ...notice Models DLS350 DLS355 Dual LED Dual LASER Source ____________________________________________________________________________________ 143 Lake St Laconia NH USA 03246 603 524 8350 Fax 603 524 8...

Page 2: ...ety page 3 1 2 Warranty page 5 1 3 Unpacking and Inspection page 5 Section 2 Feature Overview and Usage Instructions 2 1 Feature Overview page 6 2 2 Usage Instructions page 9 Section 3 Specifications...

Page 3: ...ns or provide assistance that you may require We at ODM Inc strive to provide premier customer care and technical support by providing timely responsiveness and training We are proud of our quality an...

Page 4: ...as it may cause temporary or permanent blindness or damage to the eyes Never point a laser or light source at another individual 2 Always assume that the light or laser source output is active The li...

Page 5: ...shock or damage to the instrument This includes but is not limited to a Never use an AC adapter with the instrument other than that which is specified or rated by ODM Inc b Do not use a damaged AC Ada...

Page 6: ...covered under warranty Please contact ODM Inc for instructions for appropriate RMA authorization 1 3 Unpacking and Inspection The DLS series instruments are carefully inspected and packed according t...

Page 7: ...55 are both dual wavelength single output light sources DLS350 LED source offers 850 1300 nm wavelengths for multimode testing while the DLS355 Laser Source offers 1310 1550 for singlemode application...

Page 8: ...Page 7 of 13 Figure 1...

Page 9: ...s the PWR button To turn the DLS off press PWR button again The DLS is set to turn off after 15 minutes For continuous operation press and hold the PWR button for three seconds or power on 3 Wavelengt...

Page 10: ...ed with the RP 400 series power meters The incoming 2kHz signal is identified and an audible alert is provided by the RP 400 series power meter The 2 kHz tone provided by the DLS 350 and 355 may also...

Page 11: ...040 11 Output stability 0 05dB one hour 03dB long term after 15 minute warm up Spectral Width 40nm 120nm 5nm 5nm Optical interface Universal connector contact ODM for other connector styles Tone Outpu...

Page 12: ...Adapter for Optical Light Source DLS 350 DLS 355 Cleaning Kits CK125 Cleaning Kit includes fiber cleaning solution wipes 1 25mm swabs 2 5mm swabs for 200 connectors and adapters CK250 Cleaning Kit inc...

Page 13: ...age 12 of 13 Section 5 Contact Information Customer Service and Technical Assistance Phone 603 524 8350 FAX 603 524 8332 E mail info odm inc com or support odm inc com Visit our website www ODM INC co...

Reviews: