OctiveTech 300AH Operating Instructions Manual Download Page 8

▒▓

 

www.OctiveTech.com

 Operating 

Instructions 

15

 

 
 
 
 

 

 

Fig 3.19 

 

3.3.12 System Setting 

Enter the submenu from the main menu, where it shows the system information and 
the version shown as Fig3-20 

 

Fig 3.20 

3.4 Power Supply 

The oximeter uses 4 AA Ni-MH rechargeable batteries, which can work 24 hours 
continuously.  
 
 
 
 
 
 
 
 
 
 
 

CHAPTER 4 Use Guide   

4.1 Necessary Setup before Operating 

 

              Backlight Set               
                                           
                  Always On               
         15 Second        

▒▓

 

www.OctiveTech.com

 Operating 

Instructions 

16

 

4.1.1 Adjusting date and time 

Once the oximeter being powered on, the current time will be displayed on the right top 
corner. Check whether the time is exact, if not please adjust it. As for detailed 
operation, please refer to the time setup in Chapter 3 operation guide. 

 

4.1.2 Clearing the old record

   

The measured records are not cleaned away automatically. We advise user to clear 
the previous data before operation. 
 

4.1.3 Setup of alarm limit

 

The default high/low alarm limit of the oximeter have been set when ex work, but user 
can also set the new alarm limits according to the patient’s conditions for realizing the 
best performance of oximeter.

 

As for the detailed operation, please refer to the alarm setting of Chapter 3 Operation 
Guide.

 

4.2 SPO2 (Arterial oxygen saturation) Monitor

 

4.2.1 Chief introduction

 

What is SpO2 Monitoring?   

SpO2 plethysmogram measurement is employed to determine the oxygen 
saturation of hemoglobin in the arterial blood

that is the percentage of hemoglobin 

molecules and oxygen molecules

For example, 97% hemoglobin molecules in the 

red blood cells of arterial blood combine with oxygen, then the blood has a Spo2 
oxygen saturation of 97%

the Spo2 numeric on the monitor will read 97%.The Spo2 

numeric shows the percentage of hemoglobin molecules that have combined with 
oxygen molecules to form oxhemoglobin. SpO2 /PLETH parameter can also 
provide a pulse rate signal and a plethysmogram wave.     

How do the SpO2/PLETH parameters work? 

Arterial oxygen saturation is measured by a method called pulse oximetry. It is a 
continuous, no-invasive method. It measures how much light, sent from light 
sources on one side of the sensor, is transmitted through patient tissue to a 
receiver on the other side.     

   The amount of light transmitted depends on many factors, most of which are 

constant. However, one of  these factors is blood flow in the arteries varies with 
time, because it is pulsating. By measuring the light absorbing   

during the pulsation, it is possible to get the oxygen saturation of arterial blood. 
Detecting the pulsation gives a PLETH waveform and pulse rate signal. 
Spo2 value is shown in the form of numerical value, PLETH waveform is shown in 
the form of graph on the screen.   

Summary of Contents for 300AH

Page 1: ...1 3 Electromagnetism interference 4 1 4 Conception of Warning Caution and Note 4 CHAPTER 2 OVERVIEW of SYSTEM 6 2 1 Summary 6 2 2 Specifications 6 CHAPTER 3 OPERATION GUIDE 7 3 1 Appearance 7 3 2 Disp...

Page 2: ...children Do not use the oximeter in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide Alarm must be set up according to different situation of individual patient Make su...

Page 3: ...eform SPO2 pulse column Data SPO2 PR Others Connection status of probe time and remaining battery energy Alarm Alarm SPO2 probe off Alarm mode audio alarm visual alarm flash the value and information...

Page 4: ...lay automatically If you want to power off the oximeter press the ON OFF button for several seconds 3 3 2 Function buttons Right button Enter main menu from normal display and change the setting value...

Page 5: ...eturn the main menu Attention There are three selections for the Alarm Type The first is Visual Alarm when you select this item only visual alarm is responded when alarm event happens the audio alarm...

Page 6: ...10 to display the SPO2 trend graph shown as Fig3 12 Fig 3 12 The ordinate shown as Fig 3 12 represents the value of SPO2 the abscissa represents the time 60 points can be displayed synchronously in o...

Page 7: ...nd 19 Minute 17 Hour Date Set 15 Day 12 Month 2004 Year www OctiveTech com Operating Instructions 14 Fig 3 16 Press or to highlight the item you want to set then press or to adjust the item you select...

Page 8: ...nitor 4 2 1 Chief introduction What is SpO2 Monitoring SpO2 plethysmogram measurement is employed to determine the oxygen saturation of hemoglobin in the arterial blood that is the percentage of hemog...

Page 9: ...emoglobin methylene blue Indigo carmine 4 2 3 3 Extreme illumination would affect the Spo2 measurement using nontransparent cover to shield the sensor would get the better performance Note Pulse senso...

Page 10: ...stipulations and lead to oximeter failure and possible health damage our company has right to change the warranty period a Should make the effective maintenance plan for oximeter and reused parts Thi...

Page 11: ...nce www OctiveTech com Operating Instructions 22 d Responsibility limitation During the period of warranty if user changes the parts manufactured by the third party without our company permission our...

Page 12: ...er offer oximeter Storage and Transportation Storage Storage Temperature 10 C 40 C Relative Humidity 90 Transportation Transport by air train or vessel after packing according to request Packing list...

Reviews: