21
7. TUBES
7.3.1 How to set the BIAS
Make sure that no signal is playing through the amplifier when you are adjusting the BIAS. It is
sufficient to switch on the Muting function; there is no need to disconnect the speakers.
The tubes should have reached operating temperature. The unit should run at least 15 minutes
before you adjust the BIAS. If you notice high BIAS at one or more of the output tubes in cold state
you should reduce the BIAS in any case.
Procedure
1.
Switch on the Muting function
2.
Set the bias selector switch to V1. The green LED next to the trim pot for tube V1
will light up to show that you have chosen to adjust the bias for tube V1
3.
The bias setting for this tube is correct when the green OK LED on the display illu-
minates
If the LED illuminates, the bias is set too high.
Turn the trim pot counter-clockwise until the OK LED illuminates.
If the yellow - LED illuminates, the bias is set too low.
Turn the trimmer clockwise until the OK LED illuminates.
Repeat this procedure with the other three tubes
4.
Switch the bias selector back to its "off" position.
If there is no change in the BIAS of a tube while turning the BIAS regulator, the corresponding output
tube is faulty.
Note:
It is not possible to adjust the bias if the amplifier's electronic protection has cut
in (as indicated by the red "off" LED in the display), or if the Ecomode has
powered down the unit. When you activate the measuring facility under these
circumstances, the
-
LEDs will illuminate for all tubes. You must fix the fault or
switch off the Ecomode before using the measuring facility. If a defective tube is
causing the problem, you should locate and replace it.
(See Troubleshooting
Chapter 9)
Summary of Contents for RE 320
Page 1: ...RE 320 Owner s Manual English ...
Page 33: ...34 ...