5. Water Trap:
Attach the Water Trap to the Air inlet port on the GasManifold
CAUTION: Do not attach the water trap to the outlet port of the Air Compressor as water may
condense enroute to the anesthesia machine and cause possible malfunction of some of the
components.
6. Air Hose:
Attach one end of the 15' yellow Air high-pressure hose to the inlet port of the Water Trap.
Attach the opposite end of the high pressure hose to the air compressor outlet port or to
other Air sources (wall or cylinder)
7. Waste Gas Scavenger Tubing Connection:
1. Attach the waste gas scavenger to the upper right hand side of the back frame, using the
Velcro strip or keyhole mounting, then attach the reservoir bag to bottom port of the
scavenger. Attach suction source tubing to the inlet port of scavenger control knob port.
NOTE: In military field conditions the reservoir bag to the bottom port of the scavenger
may be replaced with a large bore hose to possibly evacuate waste anesthetic out of the
work area.
.
WARNING: Scavenger system should always be mounted in order to keep debris from
entering the bellows outlet port.
1. Attach one end of the first scavenger tubing to the scavenger outlet port of the B/V
Switch labeled "to scavenger" and the opposite end to the "TEE" connector.
2. Attach one scavenger tubing, with the 19mm purple connector to the center port located
on the base of the Bellows, and attach the other end to the "TEE" connector.
3. Attach the final scavenger tubing to the top of the Waste Gas Scavenger and the other
end to the "TEE" connector.
Summary of Contents for Magellan-2200
Page 23: ...23 MAGELLAN 2200 MODEL 2 MACHINE AIR COMPRESSOR AND POWER SUPPLY BACK VIEW...
Page 26: ...26...
Page 42: ...42 GAS FLOW DIAGRAM I O2Flow from Gas Manifold to L Frame...
Page 43: ...43 GAS FLOW DIAGRAM II Air Flow from Gas Mnifold to L Frame...
Page 46: ...46 GAS FLOW DIAGRAM V Gas Flow From O2 Concentrator to O2 Flow Meter...