35
C. Pull the vaporizer forward and remove the inlet and outlet tubing connectors.
2. Mounting:
A. Reverse the preceding instructions.
CAUTION: After re-mounting the vaporizer, be certain that all four of the retaining knobs are
very secure in order to keep the vaporizer from moving during any transport or movement.
3. Utilization:
A. Pour-Fill Type: Ensure that the percent control knob is set to "0".
1. Remove filler cap.
2. Pour agent into opening, observing the "fill" chamber window to desired level.
3. Replace filler cap.
B. Key-Fill Type: Ensure that the percent control knob is set to "0"
1. Attach the keyed filler adapter to the bottle.
2. Loosen the clamp screw and remove the plug.
3. Insert keyed end of bottle adapter fully into vaporizer receiver; tighten clamp screw.
4. Raise the bottle above the filler.
5. Open the filler control latch.
6. Fill the chamber by observing the "fill" chamber window.
7. Close filler control latch.
8. Lower bottle below the level of the filler.
9. Loosen clamp screw and remove the key adapter.
10. Insert the clamp screw and tighten clamp screw.
C. Use During Procedure:
A. Turn percent control knob to desired position.
B. Adjust percent control as needed by observing the patient monitor-indicating
percentage delivered.
WARNING: Tilting the vaporizer past 45 degrees with liquid in the chamber can result in
patient injury or death.
If tilted past 45 degrees, empty the chamber, fully open the percent control, and then open
the O2 flowmeter to 5 LPM for 10 minutes.
NOTE: For complete vaporizer information, refer to the provided Penlon Sigma Delta User
Instruction Manual.
USE OF MECHANICAL VENTILATOR
Purpose: The mechanical ventilator is used to power the bellows. The air or oxygen used to
power the ventilator does NOT mix with the gas mixture that enters the inside portion of
the bellows from the flowmeters and vaporizer.
Ventilator Nomenclature:
Pneumatically Powered
Time Cycled
Inspiratory Flow Variable
Volume Constant
Square Wave Flow Pattern Only
Volume Range 6 to 14L minute volume
Pre-set High Pressure Relief Valve
@ 60 cmH
2
O
Positive End Expiratory Pressure
external (PEEP) valve
Summary of Contents for Magellan-2200
Page 23: ...23 MAGELLAN 2200 MODEL 2 MACHINE AIR COMPRESSOR AND POWER SUPPLY BACK VIEW...
Page 26: ...26...
Page 42: ...42 GAS FLOW DIAGRAM I O2Flow from Gas Manifold to L Frame...
Page 43: ...43 GAS FLOW DIAGRAM II Air Flow from Gas Mnifold to L Frame...
Page 46: ...46 GAS FLOW DIAGRAM V Gas Flow From O2 Concentrator to O2 Flow Meter...