![Oceanic DELTA 2 Service Procedure Download Page 11](http://html.mh-extra.com/html/oceanic/delta-2/delta-2_service-procedure_723230011.webp)
REGULATORS
DELTA II SECOND STAGE
© 2002 Design 1997
OCEANIC® Product Service Guide
Doc. 12-2223-r03 (10/3/10)
PG-11
R = Revision
D. Turn the locking nut further clockwise onto the poppet shaft
with small fractions of a turn, using the Poppet Installation tool
and 1/4" open end wrench. Use the correct method given in step
15 of the reassembly procedure to replace the inlet coupling after
each adjustment, and again determine whether slack is eliminated.
NOTE: Avoid tightening the locking nut any further than is
necessary to eliminate lever slack. It may be necessary to
repeat step D several times to arrive at the correct setting.
CAUTION:
Be careful to avoid over adjusting! If airflow
returns, replace the locking nut and seat with new, and start
over after rereading the above procedures.
5. Replace the diaphragm, diaphragm washer, and front cover if
removed, and pressurize the regulator again with a pure air source
of 3,000PSI. Determine the range of adjustment by performing
the following procedure:
A. Turn the adjustment knob completely out counter clockwise.
A slight to moderate airflow should be present.
B. Turn the adjustment knob completely in clockwise and fully
depress the purge button. This should initiate a slight airflow.
NOTE: If airflow is greater or less than specified for each
adjustment, refer to the troubleshooting section to determine
possible cause and treatment.
6. Purge the regulator of air, remove the In-Line Adjustment tool
and connect the LP hose directly onto the inlet coupling, using
two wrenches as prescribed in the reassembly procedure.
7. Pressurize the regulator again with a pure air source of 3,000PSI.
Return the adjustment knob to its mid range position to make a
subjective test of airflow and cracking effort. Inhale lightly through
the mouthpiece to determine that air flows easily and smoothly,
without any hesitation or lag.
NOTE: If hesitation or lag is detected, refer to the trouble-
shooting section to determine possible cause and treatment.
8. Clean and disinfect the mouthpiece in warm, soapy water before
returning to the customer.
R