Ocean Optics LS-1 Series Manual Download Page 5

 
 
LS-1 Series Tungsten Halogen Light Sources 

LS-1 Series Tungsten Halogen Light Sources    

4 

R-LS-1 Rack-mount Tungsten Halogen Light Source 

The R-LS-1 Rack-mount Tungsten Halogen Light Source is a white-light source optimized for the VIS-NIR (360-
2000 nm) range. It offers high color temperature, highly efficient output and long life in a compact, economical 
package. 

Standard R-LS-1 light sources come with a 900-hour 3100 K bulb. However, the LS-1-LL (long life) is identical to 
the LS-1 with the exception of its 10,000-hour 2800 K bulb. 

The R-LS-1 is the rack-mount version of the LS-1.  It is identical in performance to the LS-1. 

Parts Included 

The R-LS-1 light source ships with the following items: 

• 

R-LS-1 Rack-mount Tungsten Halogen Light Source 

• 

Allen wrench 

• 

Three Teflon

 

diffusion discs 

• 

One BG-34 blue filter 

• 

12 VDC power supply (a European power supply is also available) 

The Teflon

 

diffusion discs are installed inside the light source. They are used to create a diffuse light source that 

is optimal for coupling into fibers, as well as for attenuating the source when spectrometer saturation is an issue. 
From smallest to largest, the three Teflon

 

discs attenuate the light by approximately 50%, 75%, and 99% 

respectively. 

The BG-34 blue filter is a 12.7 mm filter that you can install in the light source to enhance the signal in the blue 
region relative to the red and NIR regions. 

See the Using the R-LS-1 section that follows for more information. 

Using the R-LS-1 

 

Warning 

Do NOT insert plastic or flammable materials into the 3 mm filter slot on the top of the LS-1. These 
materials could melt or ignite due to the extreme heat generated by the light source. 

The light source becomes extremely hot during operation.  Handle with extreme care. 

Connecting the R-LS-1 

Follow the instructions below to connect your R-LS-1 light source: 

1.  Identify the R-LS-1 in the equipment rack. 

2.  Connect the 12 VDC power supply to a wall outlet. 

3.  Connect the barrel connector on the power supply to the rear of the R-LS-1. 

4.  Connect a fiber to the SMA 905 connector on the R-LS-1. 

5.  Switch the lamp on using the silver On/Off switch on the R-LS-1 and allow the lamp to warm up for 

approximately 30 minutes. 

The R-LS-1 is now ready for operation in your sampling system. 

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for LS-1 Series

Page 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Page 2: ...Tungsten Halogen Light Source Standard LS 1 light sources come with a 900 hour 3100 K bulb However the LS 1 LL long life is identical to the LS 1 with the exception of its 10 000 hour 2800 K bulb Part...

Page 3: ...g ensure that the lamp has been off long enough to allow the unit to cool and disconnect the power supply from the LS 1 1 Remove the two black hex bolts located on either side of the silver On Off swi...

Page 4: ...Use the included Allen wrench to loosen the setscrew on the bottom of the lamp You do not need to remove the setscrew This screw holds the bulb in place inside the LS 1 4 Remove the two black hex bol...

Page 5: ...pectrometer saturation is an issue From smallest to largest the three Teflon discs attenuate the light by approximately 50 75 and 99 respectively The BG 34 blue filter is a 12 7 mm filter that you can...

Page 6: ...Use a 3 16 nut driver or a small adjustable wrench to remove the four rack post nuts from the top component of the rack system see diagram above Keep the four rack post nuts in a safe place 2 Lift the...

Page 7: ...f the R LS 1 card 4 Use the Allen wrench included with the R LS 1 to loosen the small silver screw It is not necessary to remove this screw completely 5 Slide the silver lamp tube out of the blue anod...

Page 8: ...lips screws on the bottom of the R LS 1 These screws hold the blue anodized bulb housing in place on the R LS 1 card 5 Slide the R LS 1 in place on the card until the recessed silver Allen screw is vi...

Page 9: ...ers contained on an S2000 spectrometer However H1 is not used by the R LS 1 It is only required when the R LS 1 is rack mounted between spectrometers The light on the R LS 1 can be toggled on and off...

Page 10: ...es Comparison of typical LS 1 Scope mode signal with and without BG 34 filter Specifications LS 1 and R LS 1 LS 1 R LS 1 Spectral range 360 2000 nm If used with an Ocean Optics spectrometer the practi...

Page 11: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Reviews: