Chapter 4
Troubleshooting
If the power supply or lamp does not seem to functioning properly, check the following:
Issue Probable
Cause
Resolution
Line power not present
Check line voltage
Power switches on, but no
LEDs light.
Fuse defective
Check fuse
Lifetime of the lamp is
exhausted
Replace the lamp module
Xenon lamp does not start
Lamp’s internal connection plug
is not closed properly
Open unit (see
Bulb Replacement
)
and close connector plug.
000-10000-080-02-0505