NORMAL TEMPERATURE REFERENCE
Oral: 35.1 - 37.3 ˚C/ 96.3 - 99.1 ˚F
Rectal:
36.2 - 37.7 ˚C/ 97.2 - 99.9 ˚F
Armpit: 35.2 - 36.7 ˚C/ 95.4 - 98.1 ˚F
Ear:
35.7 - 37.9˚C/ 96.3 - 100.2 ˚F
TECHNICAL DATA & MATERIAL
TROUBLESHOOTING
CARE AND STORAGE INSTRUCTION
Technical specifications are subject to change without notice.
Product Name:
Infrared Ear Thermometer with Backlight
Model:
MM-3600
Dimensions:
11.8 x 3.8 x 3.5 cm/ 4.65” x 1.5” x 1.38”
Weight:
40 g
Display:
Liquid Crystal Display (LCD), no mercury
Memory:
Stores 10 readings
Temperature Range:
32.0 – 43.0°C (90.0 – 109.4°F)
Accuracy:
± 0.2°C (± 0.4°F)
Operating Environment:
Temperature: 10 – 40°C (50 – 104°F)
Humidity:
≤ 95%
Storage Environment:
Temperature: -25 – 55°C (-13 – 131°F)
Humidity:
≤95%
Input Power :
One 3V lithium CR2032 battery
Approvals:
CE0197, RoHS
Applied part:
= Type B
Problem
Cause
Solution
Temperature results are not within
normal human temperature range:
32.0 to 43°C (89.6 to 109.4°F).
Lo = low; H= high.
The ambient temperature is not
within the suitable operating
temperature range:
10 – 40°C (50 – 104°F).
Place the thermometer within the suitable
temperature range: 10 – 40°C (50 – 104°F)
and try again.
Other error.
Contact local dealer.
1. Make sure the probe surface is clean.
2. Make sure the probe are correctly inserted
into ear canal.
Please contact the local dealer if the problem
continues to occur with proper operation.
- This product is not waterproof; please do not soak the entire product in water and cleaning detergent.
In the unlikely event of product malfunction, stop using immediately.
Please contact the local dealer for repair. Any tampering voids warranty.
- Clean the probe with soap and warm water or rubbing alcohol.
- Do not immerse the product into any liquid.
- Do not use any thinner or chemical solvent to clean the product.
- Keep the product out of reach and sight of children.
- Keep this product and its battery away from heat, moisture, and direct sunlight.
- Remove the batteries from the product if not used for long time.