26. pH
Probe
Calibration
1
2
3
Wash
4
Keep probe into Buffer solution
5
Calibration
Press Cal Key 3 Seconds
Set pH calibration
6
Press_cal
Calibration During 1 minutes
Wait________60s__
7
7pH_Quality_100%
Quality Probe
8
Wash
9
Keep probe into Buffer solution
10
4pH___Press_cal
Calibration During 1 minutes
Wait________60s__
11
7pH_Quality_100%
Quality Prob
12
Wash
13
14
Press Enter Key to save and exit
15
Normal Status
Note:
If you have setting Calibration = Easy the function has 1 point calibrate only 7 pH buffer solution.