background image

Adequate free 

space under see 

through face section 

for aeration, supple-

mental O2, jaw/face 

contact as necessary; prevents 

patient overheating, can more 

easily communicate with the 

patient without sound muffling.

description of components

d

es

C

rip

Tion 

of Componen

Ts

5

64878- radiolucent frame

Used to support the Torso 

support and Crescent face 

rest pad. One cam lock 

facilitates cervical flexion 

and extension.

(1) CresCent 

FaCe rest Pad

Pn em2410-sP

(1)  Contoured 

torso suPPort Pad

 

Pn 0546-06

(2)   7” x 12”  

reCtangular 

adjuster Pads

Pn 4402-06

(2)   8” x 16”  

reCtangular  

adjuster Pads

Pn 0552-06

(2)  Contoured 

torso Wedges

Pn 0550-06 (small)

Pn 0554-06 (large)  

       

large not shown

(1) radioluCent 

Frame With  

adjustable  

FaCe rest

Pn 64878

(1)  8” halF 

round 

bolster 

Pn 1581-06

(1)  8” semi 

round 

bolster

Pn 5932-06

(1)  Carry Case

Pn 64705

Cervical 

Support 

System

See Caution on page 3.

cam lock

Summary of Contents for Spine Positioning System

Page 1: ...so Support Pad 2 7 x 12 rectangular adjuster pads 2 8 x 16 rectangular adjuster pads 2 Contoured Torso Wedges pictured here with small wedge 1 Radiolucent frame with adjustable face rest 1 carry case...

Page 2: ...t to change without notice and should not be construed as a commitment by Oakworks Inc Oakworks Inc assumes no responsibility for any errors that may appear in this document nor does it make expressed...

Page 3: ...ns 2 3 Symbol Identification 4 Description of Components 5 7 Directions for Use 8 15 Maintenance Service 17 21 cleaning disinfecting 17 maintenance 18 19 list of parts 20 specifications 21 warranty 21...

Page 4: ...omfort and imaging needs for all spinal column procedures WARNING warnings A patient safety strap is required during all procedures Follow normal and required safety protocol for all procedures where...

Page 5: ...l cam locks are secure before relinquishing support of the positioning assembly The Cervical Support System see page 5 has metal parts that can cause back scatter of x rays When x rays are present wea...

Page 6: ...mbol when used in this manual or on product labels indicates that the product should be protected from moisture The humidity specifications for Transport Storage are listed on page 23 This symbol when...

Page 7: ...Crescent face rest pad One cam lock facilitates cervical flexion and extension 1 Crescent Face Rest Pad PN EM2410 SP 1 Contoured Torso Support Pad pn 0546 06 2 7 x 12 rectangular adjuster pads pn 4402...

Page 8: ...enhance patient stability by reducing sway while in the device 0550 06 small CONTOURED Torso WEDGE 0554 06 large CONTOURED Torso WEDGE The Contoured Torso Wedge is constructed of dense foam and cover...

Page 9: ...gular Adjuster pads are used either individually or combined to reduce lumbar lordosis and or increase chest height to allow for shoulders to natural ly decend out of the plane of the cervical and tho...

Page 10: ...trap is recommended during all procedures A patient saftey strap must be used during procedures where lateral roll function is used Directions for use 8 Directions for Use face rest platform adjustmen...

Page 11: ...e face rest and rotate the facerest flat against the base frame This will protect the facerest support platform during transport When placing the Spine Positioning System in the bag put some pads wedg...

Page 12: ...ch cervical procedures Components used Platform Frame 1 Crescent Face Cushion 1 Contoured Torso Support Pad 1 Small Contoured Torso Wedge 1 7 x 12 Rectangular Adjuster Pad 1 8 Semi Round Bolster 1 Dir...

Page 13: ...figured with two extra adjuster pads under the lumbar spine as a method to reduce lumbar lordosis and enhance comfort AP image of the lumbar spine Components used Platform Frame 1 Crescent Face Cushio...

Page 14: ...e T1 2 interlaminar space Contralateral oblique showing the upper thoracic facet joints Components used Platform Frame 1 Crescent Face Cushion 1 Contoured Torso Support Pad 1 Small Contoured Torso Wed...

Page 15: ...age through the C1 2 segment Lateral image through the C1 3 segments Mary AP image through the C1 2 joints Components used Platform Frame 1 Crescent Face Cushion 1 Contoured Torso Support Pad 1 Small...

Page 16: ...n Right thoracic oblique image to visualize the trajectory for a transforaminal injection Contralateral oblique showing the trajectory for targeting the mid thoracic facet joint Components used Platfo...

Page 17: ...ncluding the C7 T1 segment Directions for Use Components used Platform Frame 1 Crescent Face Cushion 1 Contoured Torso Support Pad 1 Small Contoured Torso Wedge 1 7 x 12 Rectangular Adjuster Pad 1 8 x...

Page 18: ...edures such as interlaminar epidural steroid injections facet injections medial branch blocks and medial branch radiofrequency neurotomy Components used Platform Frame 1 Crescent Face Cushion 1 Contou...

Page 19: ...Virox Disinfecting All components may be disinfected using a 10 sodium hydrochlorite bleach solution Be sure to to remove excess solution and dry thoroughly All fabric covered components may be disinf...

Page 20: ...d Replace any damaged or worn components Inspect face rest platform locking mechanism weekly Use the following procedure Step 1 Lock the platform cam Step 2 Rock platform up down Gently rock platform...

Page 21: ...hanism is to simply put a thin piece of paper between the 2 pivot parts avoiding the pins and pull on the paper If the paper pulls out cam tightening is required Maintenance Cam tightening if required...

Page 22: ...ame 1 EM2410 SP Crescent Face Cushion 1 0546 06 Contoured Torso Support Pad 1 0550 06 Small Contoured Torso Wedge 1 0554 06 Large Contoured Torso Wedge 1 4402 06 7 x 12 Rectangular Adjuster Pad 2 0552...

Page 23: ...Pad 1 5 5 x 16 x 27 0550 06 Small Contoured Torso Wedge 1 2 5 x 16 x 23 0554 06 Large Contoured Torso Wedge 1 3 5 x 16 x 29 4402 06 7 x 12 Rectangular Adjuster Pad 2 1 x 7 x 12 0552 06 8 x 16 Rectangu...

Page 24: ...rch 2010 Edition 2 May 2010 English Printed in U S A Contact Information Oakworks Medical 923 East Wellspring Road New Freedom PA 17349 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com European...

Reviews: