31

SPECIFICATIONS

Guidance and manufacturer’s declaration - electromagnetic emissions

The table is intended for use in the electromagnetic environment specified below. The customer or the 

user of the table should assure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic environment - guidance

RF emissions

CISPR 11

Group 1

The table uses RF energy only for its internal

function. Therefore, its RF emissions are very low 

and are not likely to cause any interference in 

nearby electronic equipment.

RF emissions

CISPR 11

Class A

The table is suitable for use in all establishments 

other than domestic and those directly con-

nected to the public low-voltage power supply 

network that supplies buildings used for domestic 

purposes.

Harmonic emissions

IEC 61000-3-2

Class A

Voltage fluctuations / 

flicker emissions

IEC 61000-3-3

Complies

Recommended separation distances

between portable and mobile RF communications equipment and the table

The table is intended for use in the electromagnetic environment in which radiated RF disturbances 

are controlled. the customer or the user of the table can help prevent electromagnetic interference 

by maintaining a minimum distance between portable and mobile RF communications equipment 

(transmitters) and the table as recommended below, according to the maximum output of the

communications equipment.

Rated maximum output 

power of transmitter

W

Separation distance according to frequency of transmitter (m)

150 kHz to 80 MHz

d = 1.2 √ P

80 MHz to 800 MHz

d = 1.2 √ P

800 MHz to 2.5 GHz

d = 2.3 √ P

0.01

0.12

0.12

0.23

0.1

0.38

0.37

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation 

distance d in meters (m) can be estimated using the equation applicable to the frequency of the 

transmitter, where P is maximum output power rating of the transmitter in watts (W) according to the 

transmitter manufacturer. transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

Summary of Contents for PG Series

Page 1: ...USER MANUAL www oakworksmed com 717 235 6807 ULTRASOUND SERIES made in the USA with US imported parts PG SERIES PG Series Ultrasound Series...

Page 2: ...within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc encourages requests for technical specifications and the like documen...

Page 3: ...Angle Back Support 21 Central Base Locking System 21 Locking Casters Use 21 Manual Adjustable 4 Position Leg Rest Use 22 Procedure Tray 22 Sliding Cardio Panel 23 Vascular Integrated Foot Board 23 24...

Page 4: ...ol when used in this manual or on product labels indicates a Protective Earth Ground Terminal This symbol when used in this manual or on product labels indicates that the product should be protected f...

Page 5: ...ty is 150 lbs 68 kg When lowering the table or using the Trendelenburg functions make sure there is nothing underneath the table top that can impede motion like stools cabinets accessory parts cleaner...

Page 6: ...included Table Voltage Options 120V 60Hz or 230V 50Hz Table Capacity 550 lbs 250 kg patient and accessories Widths 27 69 cm or 30 76 cm Length 3 section Fowler 34 86 cm Seat 24 61 cm Leg Rest 13 33 c...

Page 7: ...or 230V 50Hz Table Capacity 550 lbs 250 kg patient and accessories Widths 27 69 cm or 30 76 cm Length 2 section Fowler 34 86 cm Seat 38 97 cm 4 10 cm Locking Casters Locking Unlocking Casters for mobi...

Page 8: ...or 230V 50Hz Table Capacity 550 lbs 250 kg patient and accessories Widths 27 69 cm or 30 76 cm Length Rectangular Top 72 183 cm 4 10 cm Locking Casters Locking Unlocking Casters for mobile or permane...

Page 9: ...7 69 cm or 30 76 cm Length 1 section Rectangular Top 72 183 cm Length 2 section Fowler 34 86 cm Seat 38 97 cm Length 3 section Fowler 34 86 cm Seat 24 61 cm Leg Rest 13 33 cm 4 10 cm Locking Casters L...

Page 10: ...0 76 cm Length 3 section Fowler 34 86 cm Seat 24 61 cm Leg Rest 13 33 cm 4 10 cm Locking Casters Locking Unlocking Casters for mobile or permanent positioning Retractable Foot Rests Stirrups Adjustabl...

Page 11: ...ity 550 lbs 250 kg patient and accessories Widths 27 69 cm or 30 76 cm Lengths 3 section Fowler 34 86 cm Seat 24 61 cm Leg Rest 13 33 cm Adjustable Access Back Supports Left Right side Manual 4 positi...

Page 12: ...and accessories Width 30 76 cm Lengths 2 section Fowler 34 86 cm Seat 38 97 cm Adjustable Access Back Support Right Manual 4 position back support for patient positioning Manual Sliding Cardio Panel...

Page 13: ...76 cm Lengths 2 section Fowler 34 86 cm Seat 38 97 cm Integrated Foot Board Split drop away Patient Restraint Strap Must be used during all procedures Side Rails 2 Increased stability and security fo...

Page 14: ...rated Foot Board Split drop away Patient Restraint Strap Must be used during all procedures Side Rails 2 Increased stability and security for patients 4 10 cm Locking Casters Locking Unlocking Casters...

Page 15: ...is equipped with a cord having an equipment grounding conductor and a grounding pin The pin must be plugged into an appropriate outlet that is prop erly installed and grounded in accordance with all l...

Page 16: ...ional Hand Control operate all functions of the table with the touch of a button Follow the directions below to access these functions Height Down lowers tabletop Trendelenburg functions are pre serve...

Page 17: ...ssory parts cleaners etc Height Up elevates tabletop Trendelenburg functions are preserved during elevation Height Down lowers tabletop table is lowered from any position Reverse Trendelenburg activat...

Page 18: ...s again to continue optional 2 function foot control DIRECTIONS FOR USE CAUTION CONTROL OPERATIONS GENERAL Do not sit on the Fowler Backrest or Leg Rest sections Do not leave the patient unattended Wh...

Page 19: ...ain to continue Fowler Up activates fowler up to 75 Fowler Down activates fowler down to level position DIRECTIONS FOR USE CAUTION CONTROL OPERATIONS MULTI SPECIALTY WOMEN S IMAGING GENERAL WITH FOWLE...

Page 20: ...like stools cabinets accessory parts cleaners etc Height Up elevates tabletop Trendelenburg functions are preserved during elevation Height Down lowers tabletop Trendelenburg functions are preserved d...

Page 21: ...n like stools cabinets accessory parts cleaners etc Height Up elevates tabletop Trendelenburg functions are preserved during elevation Height Down lowers tabletop Trendelenburg functions are preserved...

Page 22: ...ctions make sure there is nothing that can impede motion like stools cabinets accessory parts cleaners etc Height Up elevates tabletop Trendelenburg functions are preserved during elevation Height Dow...

Page 23: ...ghtly and move to the desired position There are 3 detent positions Stop lift ing and move slightly left or right for the nearest detent position 5 To reduce stirrup length lift the handle slightly an...

Page 24: ...ot to unlock table from the floor All four casters should be locked at all times during use with a patient DIRECTIONS FOR USE CAUTION ADJUSTABLE ANGLE BACK SUPPORT Do not sit on the Adjustable Angle B...

Page 25: ...y holder Discard fluids clean and place back into tray holder CLEANING AND DISINFECTING INSTRUCTIONS If fluid spill under the seat cushion follow these instructions 1 Remove the seat pad by pulling up...

Page 26: ...L CARDIO PANEL 1 Grasp the Cardio Panel 2 Lift up slightly 3 Slide the panel to the desired opening width ELECTRIC CARDIO PANEL 1 Use the hand control or optional foot control to open the cardio panel...

Page 27: ...wler section and rotate to the 90 position Insert the rod through the paper roll and place the rod into the holes in the support arms DIRECTIONS FOR USE CAUTION VASCULAR PATIENT RESTRAINING STRAP USE...

Page 28: ...The QuickLock Face Rest can be used in both prone and supine positioning and folds out of the way when not in use Adjustments are made easily by loosening the cam locks adjusting for height and angle...

Page 29: ...roll holder support arm can be raised to the normal position This option allows the use of patient lifts The high clearance base permits patient lifts to slide easily under the center of the table fo...

Page 30: ...cquer plated metals 2 Clorox Hydrogen Peroxide Disinfectant EPA reg 67619 24 corrosion of metals deterioration of upholstery and wood lacquer leaves a hard to remove residue 3 Virox Accel TB EPA reg 7...

Page 31: ...ain wet for the disinfectant manufacturer s recommended contact time 4 Throw the used wipe s away and allow the table to dry INSPECTIONS RECOMMENDED REGULAR INSPECTIONS monthly or local standard Check...

Page 32: ...th Minimum Table Height Table Top Length Padding Comfort Plus Fabric TerraTouch Spaces not used Options Model Number Serial Number 16796025 USGNG23022723MTTZZ2S Ser USF565022 US G2 30 x 22 38 x 72 3M...

Page 33: ...urope Japan Input Service 120 VAC 15 amp 60 Hz 230 VAC 10 amp 50 60 Hz 100 VAC 10 amp 50 60 Hz Maximum Momentary Current Consumption 4 0 amps 2 0 amps 5 0 amps Voltage Output to Actuators 24 VDC 24 VD...

Page 34: ...netic environment in which radiated RF disturbances are controlled the customer or the user of the table can help prevent electromagnetic interference by maintaining a minimum distance between portabl...

Page 35: ...kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical com mercial or hospital environ ment Voltage dips short inter ruptions and voltage variat...

Page 36: ...RF transmitters as deter mined by an electromagnetic site survey ashould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the...

Page 37: ...THIS PAGE IS INTENTIONALLY LEFT BLANK 34...

Page 38: ...THIS PAGE IS INTENTIONALLY LEFT BLANK 35...

Page 39: ...THIS PAGE IS INTENTIONALLY LEFT BLANK 36...

Page 40: ...w Freedom PA 17349 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com made in the USA with US imported parts 3034177 CONFORMS TO ANSI AAMI STD ES60601 1 IEC STD 60601 1 3RD EDITION IEC STD 60601...

Reviews: