background image

Manual Part Number MMMNST0026 

Revision date: 05.14.13

1

st

 Edition,  February 2012

English,  Printed in USA

Original Instructions

I N S T R U C T I O N   M A N U A L

CONTACT INFORMATION:

O

akworks

®

 Inc.

923 East Wellspring Road 

 

 

New Freedom, PA  17349

Phone: 717-235-6807

FAX:  717-235-6798

www.oakworksmed.com

European Authorized Representative   

Emergo Europe

Molenstraat 15 

The Hague, 2513 BH Netherlands

Tel: +31 70 3458570 

Fax: +31 70 3467299 

3034177

CONFORMS TO:

ANSI/AAMI STD ES60601-1

IEC STD 60601-1 3RD EDITION

IEC STD 60601-1-2 3RD EDITION

IEC STD 60601-1-6 3RD EDITION

CERTIFIED TO CAN/CSA STD C22.2 NO. 60601-1

CB TEST CERTIFICATE AND REPORT 

PerformaLift

Series

Summary of Contents for PerformaLift Series

Page 1: ...ith optional Adjustable Side Arm Rests and Universal Table Extender Lift Assist Backrest Top Flat Top Shown with optional QuickLock Face rest platform and Boiance crescent Series www oakworks com 717...

Page 2: ...ot be construed as a commitment by OAKWORKS Inc OAKWORKS Inc assumes no responsibility for any errors that may appear in this document nor does it make expressed or implied warranty of any kind with r...

Page 3: ...10 Hand Control 11 12 Locking Casters 13 Arm Rests 14 Lowered Arm Rest 15 QuickLock Face Rest 15 Section IV Maintenance Service Cleaning 16 Inspection Maintenance 16 Troubleshooting 17 Foot and Hand...

Page 4: ...itioning capabilities for massage therapists and spa professionals of all specialties On these multifunctional professional tables the practitioner can perform numerous protocols with ease and comfort...

Page 5: ...mportant for the safety of the operator patient and any other individuals in the area of this equipment Directions for use of this equipment are described on pages 9 through 15 The operator should rea...

Page 6: ...Lus CE marked For electrical specs see Specifications Chart in the back of this manual published weight may differ due to options chosen natural material variations and padding density 3 PRODUCT DESCR...

Page 7: ...cm Length 73 185 cm Top Lift Assist Backrest Padding 3 Plush standard 4 Aero Cel or 6 Aero Cel Upholstery TerraTouch Safety Standards cETLus CE marked For electrical specs see Specifications Chart in...

Page 8: ...Padding 3 Plush standard 4 Aero Cel or 6 Aero Cel Upholstery TerraTouch Safety Standards cETLus CE marked For electrical specs see Specifications Chart in the back of this manual published weight may...

Page 9: ...manufactured for use on a nominal 120 volt circuit for the United States and a nominal 220 volt circuit for use in Europe The plug has a grounding pin see figure A in a pattern that varies by country...

Page 10: ...mine if it is level with the floor If the table rocks back and forth at all the foot that clears the floor when you rock the table will need adjustment 3 Adjust the Leveler Foot by first loosening the...

Page 11: ...orner The table weighs approximately 235 lbs 107 kg You will need 4 people working together if it is necessary to move the table Never lift the table by grasping the table top as the table top is not...

Page 12: ...ELECTRIC TOP CAUTION Do not sit on the elevated back rest or leg section Do not leave patient unattended when table is tilted near its limits WARNING When lowering the table or using the Trendelenbur...

Page 13: ...at the client take some weight off the back rest while you lift To Lower the Back Rest Press the lever under the head section and push the back rest down Release the lever and the back rest will stay...

Page 14: ...top the motion CAUTION Do not sit on the elevated back rest or leg section Do not leave patient unattended when table is tilted near its limits WARNING When lowering the table or using the Trendelenbu...

Page 15: ...12 DIRECTIONS FOR USE DIRECTIONS FOR USE HAND CONTROL OPERATIONS LIFT ASSIST BACKREST TOP FLAT TOP Table Up Head End Up Head End Down Table Down...

Page 16: ...13 DIRECTIONS FOR USE DIRECTIONS FOR USE LOCKING CASTERS LOCK To lock push locking tab down UNLOCK To unlock pull tab up as shown...

Page 17: ...be used to sustain a person s weight when getting on or off the table CAUTION Do not use the arm rest for customer weight support during the mounting or dismounting of the table The arm rest is not d...

Page 18: ...Important Safety Note Maximum Distance from edge of Table 3 CAUTION Do not sit on or put excessive weight or pressure on lowered arm shelf LOWERED ARM REST INSTRUCTIONS QUICKLOCK FACE REST INSTRUCTION...

Page 19: ...s of the table not normally cleaned on a regular basis Check for tears or cracks in the tabletop pads MAINTENANCE No specific maintenance tasks are required Oakworks medical tables are designed and bu...

Page 20: ...to be sure that it has power and that the power cable is plugged in Try each foot control and hand control to see if any work If any work then possibly there is a loose connection or a faulty controll...

Page 21: ...unction box 5 Insert new control cable and replace locking ring 6 Reconnect power cord 7 Attempt to operate the table elevation functions from the control 8 Should this fail to return the table to ope...

Page 22: ...IST OF TABLE PARTS Part Number Description 2092 1 4 Button Hand Control 55721 8 Button Hand Control 6632 4 Button Foot Control 6631 8 Button Foot Control 67721 110 Volt Control Box 67720 220 Volt Cont...

Page 23: ...ble Model Number Serial Number Part No 70537 13 rev B Power Rating US 120 V 60 Hz 4 0 A Duty Cycle MAX 2 minutes ON 18 minutes OFF Class 1 Equipment PF Series PerformaLift Tables European Authorized R...

Page 24: ...t Tabletop Applied Part Type B Applied Part Type B Applied Part Table Top IEC 529 Rating IPXO IPXO Mode of Operation Intermittent Operation MAX 2 minutes ON 18 minutes OFF Intermittent Operation MAX 2...

Page 25: ...ause 5 19 0 Recommended separation distances between portable and mobile RF communications equipment and the PERFORMALIFT The PERFORMALIFT is intended for use in an electromagnetic environment in whic...

Page 26: ...cial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital...

Page 27: ...ld strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of e...

Page 28: ...dom PA 17349 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com European Authorized Representative Emergo Europe Molenstraat 15 The Hague 2513 BH Netherlands Tel 31 70 3458570 Fax 31 70 3467299 3...

Reviews: