Service MANUAL - DTPM300 TAbLeS
Part No.:
MMMNSM0002
revisioN LeveL:
F
Page 1: ...Service MANUAL DTPM300 Tables Part No MMMNSM0002 Revision Level F...
Page 2: ...11 Hand Control Foot Control Replacement 12 13 Lift Column Replacement 14 16 Caster Replacement 17 19 Schematic 20 Wiring Diagram 21 Replacement Part s List 22 Revision History back cover Contact Info...
Page 3: ...ZZ Ser FLR561910 DTPM RT 24 x 24 40 x 78 MG TT TTCoal 01 04 2013 Order 654321 Notes Sample Product Label Model Numbers and Serial Numbers always start with a letter The model number and serial number...
Page 4: ...epeat using the foot control Check for abnormal noises Replace any missing or illegible labels Check that all fasteners are present and fastened securely Check table grounding Clean unusual buildup of...
Page 5: ...le manufactured after 02 25 13 3 See Control Box Replacement If the hand control and or foot control are not working 1 Defective Hand Foot Control 1 See Hand Foot Control Replacement If one or both of...
Page 6: ...below 4 Check all connections on the control box of the table See photo below 5 Make sure that all male and female connectors are firmly pushed into each other Even a slight misconnection can cause po...
Page 7: ...e buttons when the beeping stops 2nd row buttons Down button 2 Run all the motors in the system to the fully retracted position Down Caution Both columns should move down unless already fully down If...
Page 8: ...imes it takes sev eral attempts to press them simultaneously The control box will beep approximately 19 times You must press buttons to level or retract columns within 10 seconds or it will exit manua...
Page 9: ...ks for further assistance if your table does not reset 717 235 6807 Down button 3 Run all the motors in the system to the fully retracted position Down Caution If the tilt on the table starts to Incre...
Page 10: ...ximately 10 times Release the buttons when the beeping stops Top row buttons Down button 2 Run all the motors in the system to the fully retracted position Down Caution If the tilt on the table starts...
Page 11: ...source Lastly give yourself adequate room around the table to make the task more comfortable and safe PLEASE READ INSTRUCTIONS COMPLETELY BEFORE REPAIR 1 Identify and mark cables Masking tape works g...
Page 12: ...this to insure control box has power Photo 1 Activated Photo 2 Depressed 3 Remove the strain relief and all cable from the control box Remove ground wire from control box and disconnect power cord usi...
Page 13: ...ut of the way Make sure all cables are not in a pinch area 8 It may be necessary to reset the control box See table resetting a TABLE RESET TABLES MANUFACTURED BEFORE 03 15 12 b TABLE RESET TABLES MAN...
Page 14: ...CTIONS COMPLETELY BEFORE REPAIR 1 Hand Control Replacement Remove Strain Relief Plug the new Hand Control into the Control Box Test new Hand Control and if this does not work see Control Box Replaceme...
Page 15: ...control box See table resetting a TABLE RESET TABLES MANUFACTURED BEFORE 03 15 12 b TABLE RESET TABLES MANUFACTURED 03 15 12 02 25 13 c TABLE RESET TABLES MANUFACTURED AFTER 02 25 13 4 Please fill out...
Page 16: ...stly give yourself adequate room around the table to make the task more comfortable and safe PLEASE READ INSTRUCTIONS COMPLETELY BEFORE REPAIR It will be necessary to power the table for varies tasks...
Page 17: ...functioning column end of the table 3 2 people required Carefully roll the top and lift off of the main bearings then roll the rest of the way till the top lifts off completely The top is heavy and aw...
Page 18: ...thout lifting the base from the floor and above the base for the strain relief and power cord removal 6 Reassemble in the reverse order Torque bolts in column bottom and top to 15 Ft lbs 7 Slide the t...
Page 19: ...ure that the table is clear of all people and materials before attempting to repair Remove the power cable from its power source Lastly give your self adequate room around the table to make the task m...
Page 20: ...lding the caster to keep it from rotating 4 Raise the base and insert a wood spacer underneath so the caster is not touching the ground 5 Unscrew the locked caster It should break free easily with han...
Page 21: ...ighten with a socket wrench turn past flush 8 Using a general purpose cement glue down the black rubber plastic cover We use Loctite super glue gel Do not get gel on the screw threads 9 Please fill ou...
Page 22: ...EDGES UNLESS OTHERWISE NOTED ALL DIMENSIONS ARE IN INCHES X X 03 X XX 02 X XXX 005 ANGLE 1 DO NOT SCALE DRAWING AUTHOR OAKWORKS INC NOTES DESCRIPTION REVISIONS REV DATE MATERIAL DTPM 300 SCHEMATIC DTP...
Page 23: ...21 Wiring diagram Table of Contents...
Page 24: ...Control box 230V Systems 1 57916 2 Control Box CB16 220V Openbus Hand Control 1 Each 57920 57920 1 Hand Control 4 Button Hand Control Overlay 4 Button DTPM Foot Control 1 4027 Foot Control 2 Button L...
Page 25: ...tion IEC STD 60601 1 6 3rd Edition Certified to CAN CSA STD C22 2 No 60601 1 CB TEST CERTIFICATE AND REPORT Service MANUAL dtpm300 Tables Part No MMMNSM0002 Revision Level F Revision History Revision...