INSTALLATION

The Tables come completely assembled and ready to use. Plug the cord 

into a functioning outlet that is rated for the table.

Arrange the power cord and control cords so that they will not create a 

tripping hazard and where the controls are located to your liking and 

are conveniently accessible.

Be sure access to plug is not blocked for disconnecting the table from 

power.

This product must be grounded. If it should malfunction or break down, 

grounding provides a path of least resistance for electrical current to reduce 

the risk of electric shock. This product is equipped with a cord having an 

equipment-grounding conductor and a grounding plug. The plug must be 

plugged into an appropriate outlet that is properly installed and grounded in 

accordance with all local codes and ordinances. See U.S. sample below.

This product is manufactured for use on a nominal 120-volt circuit for the 

United States and a nominal 220-volt circuit for use in Europe. The plug has 

a grounding pin (see figure A) in a pattern that varies by country as shown 

in the Product Identification section of this manual. Make sure that the 

product is connected to an outlet having the same configuration as the plug.

Correct Implementation

Incorrect Implementation

DANGER-

Improper connection of the equipment-grounding conductor 

(see figure B) can result in a risk of electric shock. Check with a quali-

fied  electrician or service man if you are in doubt as to whether the product 

is properly grounded. Do not modify the plug provided with the product 

- if it will not fit the outlet, have a proper outlet installed by a qualified 

electrician.

To reduce the risk of electric shock, this furnishing has a polarized plug. 

This plug will fit in a polarized outlet only one way. Do not change the 

plug in any way.

 

Be sure to read all 

 

 

cautions, warnings    

 

and instructions given   

in the manual to  prevent 
injury to both operator  
and client.

GROUNDING

INSTALLATION

WARNING

Figure A

Figure B

DANGER

Summary of Contents for PG Series

Page 1: ...INSTRUCTION MAINTENANCE MANUAL www oakworksmed com 717 235 6807 PG300 PG200 PG350 made in the USA with US imported parts Oakworks PG Series...

Page 2: ...nd should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc assumes no responsibility for any errors that may appear in this document nor does it make expressed or implied warranty of any...

Page 3: ...4 Position Leg Section 11 Paper Roll Holder 12 Table Extender 12 QuickLock Face Rest 13 Patient Lift Compatible Base 14 T Rail Adapter 14 Detachable Arm Board 14 Section IV Maintenance Service Cleani...

Page 4: ...ard Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual Portable and mobile RF...

Page 5: ...table or using the Trendelenburg Reverse Trendelenburg function make sure there is nothing underneath the table top that can impede motion like stools cabinets accessory parts cleaners etc Use this t...

Page 6: ...rs Heavy duty 4 Locking Paper Roll Holder Accommodates 21 53cm roll roll not included TableWeight 359 lbs 163kg Warranty 3 years parts 1 year labor Safety Listings FDA listed cETLus CE marked OPTIONS...

Page 7: ...Roll Holder Accommodates 21 53cm roll roll not included TableWeight 359 lbs 163kg Warranty 3 years parts 1 year labor Safety Listings FDA listed cETLus CE marked OPTIONS CODE Central Base Locking Sys...

Page 8: ...per Roll Holder Accommodates 21 53cm roll roll not included TableWeight 359 lbs 163kg Warranty 3 years parts 1 year labor Safety Listings FDA listed cETLus CE marked OPTIONS CODE Central Base Locking...

Page 9: ...olt circuit for the United States and a nominal 220 volt circuit for use in Europe The plug has a grounding pin see figure A in a pattern that varies by country as shown in the Product Identification...

Page 10: ...Table DOWN Control lowers tabletop table is lowered from any position CAUTION Do not leave clients unattended when table is tilted near its limits 1 Head End Tilt Control activates tilt up to 15 posit...

Page 11: ...activates tilt up to 15 positioning stops when table is level press again for complete reverse Backrest activates backrest up to 75 Backrest activates backrest down to level position CAUTION Do not si...

Page 12: ...manual adjustable foot rest to allow access to the stirrup 3 Fold out to fully open position 2 Grasp Stirrup handle pull out firmly while lifting slightly Pull fully out 4 To rotate stirrup to the de...

Page 13: ...OCK Lift up on locking arm with your foot to unlock table from the floor CENTRAL BASE LOCKING SYSTEM The table base can be locked in two ways to prevent caster movement The easiest way is to use the c...

Page 14: ...dure Use a slow or medium speed If you raise the leg section too fast the height locking mechanism will not engage The leg section will lock into the closest height position TO LOWER THE LEG SECTION R...

Page 15: ...ounting Plate PAPER HOLDER USE To load the paper pull down on two paper holder support arms located on the under side of Fowler Rotate to 90 position Insert rod through paper roll and place rod into t...

Page 16: ...replacing the QuickLock Face Rest just push the frame back into the outlets being sure to keep the dowels within 3 of the table to ensure a stable connection For complete use care instructions see pro...

Page 17: ...ght or pressure on arm board NOTE on 27 wide tables the paper roll holder support arm must be folded down to allow installation of the T Rail adapter next to the paper roll holder After installation t...

Page 18: ...s of the table not normally cleaned on a regular basis Check for tears or cracks in the tabletop pads MAINTENANCE No specific maintenance tasks are required Oakworks medical tables are designed and bu...

Page 19: ...nnection or a faulty controller It may be necessary to do a reset of the control box a Press 2 buttons shown here simultaneously on foot control sometimes it takes several attempts to press them simul...

Page 20: ...the table functions See troubleshooting section for details To replace the foot control follow these steps IMPORTANT Always unplug power cord before removing or connecting cables to the control box SA...

Page 21: ...Box U S 57916 220 Control Box European 57410 Lifting Column 57917 Back Rest 61286 Paper Roll Holder Rod 61736 06 Paper Hold Down Strap LIST OF TABLE PARTS View complete warranty details at www oakwor...

Page 22: ...ass 1 Equipment Dual Tower Ultrasound Tables European Authorized Representative Emergo Europe Molenstraat 15 The Hague 2513 BH Netherlands 923 East Wellspring Rd New Freedom PA 17349 USA Phone 717 235...

Page 23: ...ry Current Consumption 4 0 amps 2 0 amps 5 0 amps Voltage Output to Actuators 24 VDC 24 VDC 24 VDC Electric Shock Protection Class 1 Equipment Class 1 Equipment Class 1 Equipment Tabletop Applied Part...

Page 24: ...RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class A The Ultrasoun...

Page 25: ...tial mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61...

Page 26: ...ers as deter mined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following...

Page 27: ...24 NOTES NOTES...

Page 28: ...6807 FAX 717 235 6798 www oakworksmed com European Authorized Representative Emergo Europe Molenstraat 15 The Hague 2513 BH Netherlands Tel 31 70 3458570 Fax 31 70 3467299 made in the USA with US impo...

Reviews: