8. Probe Care and Maintenance
Keep the conductivity probe clean. Rinse the probe twice,
and gently swirl it while you take readings. For best accu-
racy, soak a dry probe for at least 5 to 10 minutes or
longer before calibration. Wash the probe with deionized
or tap water before storing it. Never scratch the platinum
portions with a hard substance. Do not strike the probe
against any hard surface.
Do not make continuous contact with your solutions.
Readings will rise over a period of time while you soak
your probe.
Do not immerse the probe in oily solutions. Clean the
probe thoroughly by stirring it in mild detergent bath.
Wipe the probe with a soft tissue paper. Wash thoroughly
in tap water and then in deionized water. Recalibrate the
meter after cleaning the probe.