Instruction Manual

 

 

pH 5+ 

pH/°C 

pH 6+ 

pH/°C/mV 

Ion 6+ 

pH/°C/mV/Ion 

 

 

 

 

 

Technology Made Easy ...

 

      

 

 

68X243633 Rev 

1-1

  

3

/201

1

 

Part of Thermo Fisher Scientific 

Summary of Contents for pH 5+

Page 1: ...Instruction Manual pH 5 pH C pH 6 pH C mV Ion 6 pH C mV Ion Technology Made Easy 68X243633 Rev 1 1 3 2011 Part of Thermo Fisher Scientific...

Page 2: ......

Page 3: ...as a step by step guide to help you operate the meter second it serves as a handy reference guide This manual is written to cover as many anticipated applications of the pH 5 pH6 and Ion 6 meters as...

Page 4: ...ion 7 3 1 1 pH Calibration Procedure 7 3 1 2 Changing the pH Buffer Group 8 3 1 3 Resetting User Calibrated Values 9 3 2 Ion Calibration Ion 6 9 3 2 1 Ion Calibration Procedure 10 3 3 Millivolt mV Cal...

Page 5: ...and temperature o C The pH 6 and Ion 6 meters measure pH mV ORP and temperature Additionally the Ion 6 allows direct ion concentration measurement of various ions mono and divalent The mV mode is als...

Page 6: ...ter starts up in the mode that you last switched off from MODE INC Selects measurement mode for Ion mV pH and Temperature Increment button for mV calibration pH 6 only CAL Allows calibration for Ion p...

Page 7: ...y condition 2 3 Inserting Removing the Rubber Armor Stand 1 To remove meter from rubber armor push out from the bottom edges of meter until it is completely out of boot Ensure that cables of ISE pH el...

Page 8: ...until it locks 2 5 Battery Replacement The LO annunciator of the LCD alerts you when battery power is running low Caution Power off the meter before changing battery 2 6 Connecting the Electrode and...

Page 9: ...ey All LCD segments will display momentarily as the meter performs a self diagnostic test The Ion 6 will display if the meter has not been calibrated or if the meter has been reset 2 Press MODE key to...

Page 10: ...ue will not be accepted Press CAL to abort calibration and resume measurement Always use new pH buffer solutions for calibration Do not reuse buffer solutions as it may be contaminated and affect the...

Page 11: ...everting to pH measurement mode 6 For highest accuracy perform a multiple point calibration Repeat step 1 with additional pH buffer calibration standard solutions 3 1 2 Changing the pH Buffer Group Yo...

Page 12: ...ration Ion 6 The Ion 6 meter is capable of 2 or 3 point ion calibration with standard solutions The Ion 6 will display if the meter has not been calibrated or if the meter has been reset To abort ion...

Page 13: ...h calibration or sample measurement to avoid cross contamination 3 2 1 Ion Calibration Procedure The Ion 6 can measure various ions Ion measurement requires ion selective electrodes ISE sold separatel...

Page 14: ...ndard solution Add ISA if required Swirl it gently Press CAL key to begin calibration mode 6 Press ENTER to confirm that the ppm value is the desired standard The displayed value now shows the corresp...

Page 15: ...ss MODE key to enter mV mode the LCD displays mV 2 Dip the ORP electrode into a known standard solution e g Quinhydrone 255 and swirl it until the reading stabilizes 3 Press CAL key to enter mV calibr...

Page 16: ...displayed value to a NIST certified thermometer or other thermometer known to be accurate For best accuracy place both the probe and thermometer in a constant temperature bath 3 Press CAL key to enter...

Page 17: ...o a constant temperature bath 3 Press CAL key to enter temperature calibration mode The LCD shows CA momentarily and displayed reading flashes Note that this displayed value should either be 25 0 o C...

Page 18: ...ection 4 3 4 Rinse probes with clean water before taking next reading or storage 4 2 Millivolt mV Reference Check Ion 6 only The mV mode in Ion 6 can be used for the diagnosis of ISE or pH electrode c...

Page 19: ...have 2 point calibration Perform 2 or 3 point ion calibration LO displays in the LCD Low battery Replace batteries Unstable reading a Electrode not deep enough in sample b Dirty electrode c Broken ele...

Page 20: ...0 10 01 USA pH 4 01 6 86 9 18 NIST pH 4 10 6 97 Pb 3 3 3 Temperature Range 0 0 to 100 0 o C 3 3 3 Resolution 0 1 o C 3 3 3 Accuracy 0 5 o C 3 3 3 Temperature Comp Automatic Manual 0 to 100 o C 3 3 3 M...

Page 21: ...rying case ECPH601PLUSK 01X245028 pH 6 with pH and ATC probes and solutions in hard carrying case ECPH602PLUSK 01X245026 pH 6 with pH ATC probe and solutions in hard carrying case ECPH603PLUSK 01X2450...

Page 22: ...e 1 pint ECBU4BT 00654 00 pH 4 01 buffer sachets 20 mL x 20 pcs ECBU4BS 35653 01 pH 6 86 buffer solution 480 mL bottle ECBU686BT 00654 03 pH 7 00 buffer solution 480 mL bottle 1 pint ECBU7BT 00654 04...

Page 23: ...or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products 10 RETURN OF ITEMS Authorization must be obtained from ou...

Page 24: ......

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ...Oakton Instruments 625 E Bunker Court Vernon Hills IL 60061 USA Tel 1 888 462 5866 Fax 1 847 247 2984 info 4oakton com www 4oakton com Eutech Instruments Pte Ltd Blk 55 Ayer Rajah Crescent 04 16 24 S...