Sensor Replacement

Screw off the sensor ring, unplug the sensor, plug in the new 

replacement sensor (pay attention to the probe’s position), and 

rescrew on the sensor ring. 

Battery Installation

The tester uses four AAA batteries. Please install batteries according to 

the following steps. Note the correct direction of battery installation: the 

positive side (+) of every single battery must face up. Incorrect installa-

tion of batteries will cause damage to the tester and create a potential 

hazard. 

1.  Pull the battery cap up.
2. Slide the battery cap along the direction of arrow.
3.③Open the battery cap.
4. Insert the batteries 

(ALL POSITIVE SIDES FACING UP).

 

5. Close the battery cap.
6. Slide and lock the battery cap along the direction of arrow.
7.  Fit the tester’s cap while making sure to push all the way 

down. The  tester’s waterproof design may be compromised if 

the cap is not fitted correctly.

4 x AAA Batteries

+

+

+

+

_ _

_ _

1

2

3

4

5

6

7

Warranty

We warrant this instrument to be free from defects in material and 

workmanship and agrees to repair or replace free of charge, at 

option of Oakton Instruments, any malfunctioned or damaged prod-

uct attributable to responsibility of Oakton Instruments, for a period 

of 

two years

 from the delivery (a 

six-month

 limited warranty applies 

to sensors). This warranty does not apply to defects resulting from 

actions such as misuse (violation of the instructions in this manual 

or operations in the manner not specified in this manual), improper 

maintenance, and unauthorized repairs. Warranty period is the time 

limit to provide free service for the products purchased by custom-

ers, not the service life of the tester or probe.

Oakton Instruments reserves the right to update the information in 

this manual without giving notice in advance. 

Self-Diagnostic Messages

Symbol

Self-diagnostic  

information

How to fix

Er 1

Wrong conductivity 

standard solution, which 

exceeds the recognizable 

range of the meter.

1.  Check if standard solution is 

correct.

2. Check if electrode is damaged.

Er 2

CAL/

 

 is pressed before 

measurement is stable  

(smiley face 

(

)

 appears).

Wait for the measurement stability 

icon 

(

)

 to appear and stay, then 

press 

CAL/

 

  key.

Specifications

Conductivity

Range

0 to 200.0 μS/cm, 0 to 2000 μS/cm,  

0 to 20.00 mS/cm

Resolution

0.1 μS/cm, 1 μS/cm, 0.01 mS/cm

Accuracy

±1% full-scale

Calibration

1, 2, or 3 points; automatic recognition of 

standards

TDS

Range

0 to 100.0 ppm, 0 to 1000 ppm,  

0 to 10.00 ppt

Resolution

0.1 ppm, 1 ppm, 0.01 ppt

Salinity

Range

0 to 10.00 ppt

Resolution

0.01 ppt

Temperature

Range

32 to 122°F (0 to 50°C)

Resolution

0.1°F/°C

Accuracy

±0.9°F (0.5°C)

 

Display:

 LCD with two-color backlight. Blue = measurement mode; 

Green = calibration mode

Reading lock:

 HOLD icon

Power:

 four AAA batteries (included); >400 hours of continuous 

operation

Low-voltage warning:

 

 battery status icon flashes

Auto power-off:

 tester automatically turns off after 8 minutes  

of nonuse

IP rating:

 IP67 (waterproof), floats on water when sensor cap is on

Dimensions (L x W x H):

 7" x 1.5" x 1. 5" (17.8 x 4 x 4 cm)

Weight:

 4.7 oz (133 g)

Ordering Information

Model

Product description

Catalog  

number

CTS5

CTSTestr 5 pocket multiparameter tester

35634-42

Replacement sensor for CTSTestr 5

35634-44