background image

 

CTSTestr

 5 Pocket Tester       

Features

•  Measuring parameters:  

conductivity, TDS, salinity, 

temperature

•  Large easy-to-read LCD with 

2-color backlight

•  Replaceable single-junction 

sensor saves you money

•  Auto power-off function  

conserves battery life

• IP67 waterproof rating

Operating Manual

 

Thank you for purchasing the Oakton

®

 CTSTestr

 5 Multiparameter 

Pocket Tester. Please carefully read this operating manual before 

using the product to obtain an accurate and reliable test result  

and avoid unnecessary damage to the tester or probe.

Conductivity Measurement

1.  Press  /MEAS key to turn on the tester. Rinse the probe in  

distilled water and dry it.

2.  Dip the probe in sample solution, stir gently, and allow it to  

stand still in the solution until a stable reading is reached and  

the stability indicator (

) remains on, then take a conductivity 

reading.

3.  Press MODE/

 key to switch from Conductivity to TDS to Salinity 

parameters.

Notes

•  The TDS and salinity measurements are converted from the  

conductivity measurements via a certain conversion factor.

•  The tester can calibrate 84 μS/cm, 1413 μS/cm, and 12.88 mS/cm  

conductivity calibration solution. User can conduct 1 to 3 point  

calibrations (see table below). Usually calibrating the tester with  

1413 μS/cm solution alone meet testing requirements.

Calibration 

indication icon

Calibration 

standard

Measuring 

range

L

  84 μS/cm

0 to 200 μS/cm

M

  1413 μS/cm

200 to 2000 μS/cm

H

12.88 mS/cm

2 to 20 mS/cm

•  The tester has been calibrated before leaving the factory. Generally, 

users can use the tester directly or users can test conductivity buffer 

solutions first. If the error is large, then calibration is needed.

•  For conductivity calibration solutions, we recommend that users 

replace with new solution after each calibration to keep the  

standard solution’s accuracy. Do NOT pour the used calibration 

solution back into the solution bottle in case of contamination.

•  For temperature compensation factor, the default setting of the 

temperature compensation factor is 2.0%/°C. User can adjust the 

factor based on test solution and experimental data in parameter 

setting P3. 

Solution

Temperature compensation factor

NaCl

2.12%/°C

5% NaOH

1.72%/°C

Dilute ammonia

1.88%/°C

10% Hydrochloric acid

1.32%/°C

5% Sulfuric acid

0.96%/°C

•  TDS and conductivity are linear related, and the conversion factor 

is 0.40 to 1.00. Adjust the factor in parameter setting P4 based 
on the requirements in different industries. The factory default 
setting is 0.71. Salinity and conductivity are linear related, and the 
conversion factor is 0.5. The tester only needs to be calibrated in 
Conductivity mode, then after calibration of conductivity, the tester 
can switch from conductivity to TDS or salinity. 

• 1000 μS/cm = 1 mS/cm; 1000 ppm = 1 ppt

•  

Conversion Example:

 If conductivity measurement is 1000 μS/cm

2

then the default TDS measurement will be 710 ppm (under the 
default 0.71 conversion factor), and the salinity be 0.5 ppt.

Setting the Parameters

When tester is off, long press  /MEAS key to enter parameter set-

ting mode. Short press MODE/

 key to switch from P1 to P2…P7. 

Short press CAL/

  key and parameter will flash, then short press 

MODE/

 key to choose desired parameter. Short press CAL/

  key to 

confirm selection. Long press  /MEAS key to exit parameter setting 

mode.

Symbol

Menu setting

Selection

Factory default

P1

Auto lock (HOLD)

Off – On

Off

P2

Backlight

Off – 1 – On

1 (1-min auto-off)

P3

Temperature 

compensation factor

0.00 to 4.00%

2.00%

P4

TDS factor

0.40 to 1.00

0.71

P5

Salinity unit

ppt – mg/L

ppt

P6

Temperature unit

°C – °F

°C

P7

Restore to factory 

default

No – Yes

No

Notes

•  For Automatic Lock (P1), select “On” to activate auto lock function. 

When reading is stable for more than 10 seconds, the tester will 

lock the value automatically, and “

HOLD

” icon will appear on the 

bottom left of the screen. Press CAL/

  key to cancel HOLD on  

reading.

•  For Backlight (P2), select “Off” to turn off backlight function,  

“On” to turn on backlight function, or “1” to have backlight last  

for 1 minute.

•  For Factory Default (P7), select “Yes” to restore the calibration to 

theoretical values and parameter settings to initial values. When 

tester’s calibration or measurement performs abnormally, this  

function can be adopted so the tester returns to factory default  

setting and then users can conduct calibration or take measure-

ments again.  

Keypad Functions

Short press

 = <2 seconds         

Long press

 = >2 seconds

MEAS

1.  Short press to turn on the tester and long press to turn off the 

tester.

2.  When turned off, long press to enter parameter setting mode. 

3.  In measurement mode, short press to turn on backlight.

MODE

1.   In measurement mode, short press to switch from 

COND

  

to 

TDS

 to 

SAL

.

2.  In parameter setting mode, short press to change parameter 

(unidirectional).

CAL

<

1. Long press to enter calibration mode.

2. In calibration mode, short press to confirm calibration.

3.  When measured value is locked (HOLD icon), short press to 

unlock.

Conductivity Calibration

1.  Press MODE/

 key to switch to conductivity measurement mode. 

Rinse the probe in distilled water and dry it.

2.  Pour a small amount of 1413 μS/cm and 12.88 mS/cm conductivity 

calibration solution into calibration bottles.

3.  Long press CAL/

  key to enter calibration mode, short press  

/MEAS key to exit.

4.  Dip the probe in 1413 μS/cm conductivity calibration solution,  

stir gently and allow it to stand still in the solution until a stable 

reading is reached. When stability icon (

) appears and remains 

on the screen, short press CAL/

  key to complete the one-point 

calibration. The tester returns to measurement mode and  

indication icon “

M

” will appear at the bottom left of the screen.

5.  For a 2-point calibration. After 1st point calibration, dip the probe 

in 12.88 mS/cm conductivity calibration solution. If the value is 

accurate, it is not necessary to conduct a 2nd point calibration. If  

it is inaccurate, repeat steps #3 and #4 above to complete the  

2nd point of calibration using 12.88 mS/cm buffer solution. 