Instruction Manual
CON 11/110
42
8.3
P2.0: Viewing calibration data
This mode lets you recall the previous calibration data on each of the five measuring
ranges. This is a “view only” mode.
From measurement mode:
1. Press the SETUP key to enter Set Up menu.
2. Press the MI/
▲
or MR/
▼
key until you view
parameter P2.0.
3. Press
the
HOLD/ENTER
key repeatedly to view the
previous calibration data on each of the five
measuring ranges.
4. Once you have scrolled through all calibration data,
you will automatically return to the subgroup menu.
5. Press CAL/MEAS key to return to the measurement
mode.
NOTES: If there is no previous calibration data at a particular
point, the primary display will show “----“.
SETUP
cal
84.0
p 2.0
SETUP
---
p 2.1
SETUP
p 2.2
HOLD
ENTER
HOLD
ENTER
µ
S
Figure 35: P2.0 –Viewing
previous calibration data
for each measurement
range