background image

n

  

pH Calibration

For best results, periodic calibration with known accurate standards is recommended. The 700 

series meters can be calibrated with up to 5 buffers. The non-volatile memory retains all calibration 

values upon meter shut down. 

The following calibration standards are automatically recognized; 

To eliminate temperature errors associated with the pH electrode, attach the automatic 

temperature compensation (ATC) probe for best accuracy. 

1. Press 

 as needed to select pH.

2.  Dip the pH and ATC electrodes into pH buffer and press 

. The secondary display will lock on  

  the appropriate buffer value. Provide stirring for best results. 

  When the 

READY

 indicator appears, press 

 to accept. The primary reading will flash briefly   

  before the secondary display begins scrolling the remaining available buffers.

3.  Rinse the pH and ATC electrodes then dip into the next pH buffer. The secondary display will  
  lock on the appropriate buffer value. When the 

READY

 indicator appears, press 

 to accept.  

  The primary reading will flash briefly then display the percent efficiency (slope) before the  
  secondary display begins scrolling the remaining available buffers.

4.   To calibrate another buffer repeat step 3 or press 

 to return to the measurement mode.

n

  

Temperature Calibration

The thermistor sensor used for automatic temperature compensation and measurement is  

both accurate and stable. Temperature calibration is recommended upon electrode  

replacement, whenever the temperature reading is suspect, or if matching against a certified 

thermometer is desired. 

1.  Connect the temperature probe to the meter and place into a solution with a known accurate 
   temperature such as a constant temperature bath.

 

Note:

 To adjust the manual temperature compensation (MTC) value, do not connect the  

 

  temperature probe. 

2. Press 

 as needed to select pH or mV/R.mV.

3. Press 

 followed by 

.  The primary display shows the measured temperature while the  

  secondary display shows the factory default temperature.

4.  Adjust the temperature using 

 or 

. Press 

 to accept or 

 to cancel. The meter  

  allows an adjustable maximum value of ± 5 

º

C (or ± 0.9 

º

F) from the factory default temperature. 

 

n

  

Millivolt (mV) Offset Adjustment

Oxidization Reduction Potential (ORP or Redox) is not a precise measurement, but is useful as a 

relative indicator. mV offset adjustment makes readings comparable to a reference.

1.  Connect an ORP electrode and press 

 as needed to select mV (or R.mV).

2.  Dip the ORP electrode into a solution with a known mV value and stir. 

3. Press 

 when the reading is stable. The primary display shows the relative millivolt value  

  (R.mV) while the secondary display shows the factory default mV value.

4.  Adjust the R.mV value using 

 or 

. Press 

 to accept or 

 to cancel.  

  The meter allows an adjustable maximum value of ± 150 mV from the factory default  
  mV value. 

 

Note:

 

When an offset has been stored successfully, R.mV replaces mV. 

n

  Meter Connections

pH

BNC connection for pH, ORP (Redox), or  

Ion Selective Electrodes (ISE).

REF

Pin connection for half cell reference electrodes. 

Requires separate half cell BNC electrode.

Note: REF is not commonly used and is not required.

ATC

For Automatic Temperature Compensation probe.

DC

Power supply.

CALIBRATION

USA buffer group

1.68, 4.01, 7.00, 10.01, 12.45

NIST buffer group

1.68, 4.01, 6.86, 9.18, 12.45

DC

ATC REF.

pH

3

2

Reviews: