6. HOLD function

This feature lets you freeze the value of the
pH or conductivity reading for a delayed
observation. HOLD can be used any time
when in MEAS mode.

1. 

To hold a measurement, press the
HOLD key while in measurement
mode. "HOLD" will appear on the 
display. 

See Figure

2. 

To release the held value, press HOLD
again. Continue to take measurements.

NOTE:

This meter will hold a reading 
for up to 30 minutes, because it 
features automatic shutoff after 
30 minutes to conserve batteries. 

pH

ATC

MEAS

HOLD

C

A

A

5. Measurement

The READY indicator appears on the display when the readings stabilize. The read-
ing holds until the measured value exceeds the tolerance, then the READY annuncia-
tor turns off.

NOTE:

Be sure to remove the protective
rubber cap of the electrode before
measurement.

To take readings:

1. 

Rinse the probe with deionized or 
distilled water before use to remove any
impurities adhering to the probe body.
If the pH electrode has dehydrated,
soak it for 30 minutes in OAKTON 
electrode storage solution or a 2M–4M
KCl solution.

2. 

Switch on the meter. The MEAS annun-
ciator appears on the top center of the
LCD. The ATC indicator appears in the
lower right hand corner to indicate
Automatic Temperature Compensation.
Press MODE to select desired measur-
ing parameter (pH or conductivity).

See Figure 

3. 

Dip the probe into the sample.

NOTE: 

When dipping the probe into the
sample, make sure the tip of the
probe is completely immersed. Stir
the probe gently in the sample to
create a homogenous sample. Be
sure to tap probe to remove air bubbles.
Air bubbles will cause errors in the
reading.

4. 

Allow time for the reading to stabilize.
Note the reading on the display. When
the reading is stable, the READY
annunciator appears.

5. 

To toggle between pH and conductivity
readings, press the MODE key.

See figures        and 

NOTE: 

Conductivity readings are auto-

ranging and will automatically
move to the correct range (four
ranges possible). 

ATC

READY

MEAS

µ

  S

C

C

pH

ATC

READY

MEAS

C

B

pH

ATC

MEAS

C

A

A

pH measurement mode display

Conductivity measurement mode display

C

B

16

Summary of Contents for 35630-00

Page 1: ...will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Au...

Page 2: ...ting 19 9 Error Messages 20 10 Specifications 21 11 Accessories 22 23 12 Warranty 24 13 Return of Items 24 2 3 pH Conductivity C Meter HOLD MODE ON OFF CAL MEAS C MEAS pH ATC pH Con 10 Series ENTER 1...

Page 3: ...nding graphic indicator on the LCD ON OFF Powers and shuts off the meter HOLD Freezes the measured reading To activate press HOLD while in measurement mode To release press HOLD again MODE Selects the...

Page 4: ...s a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes or electrodes For a replacement probe see the Accessories section page 22 NOTE Keep connector dry and c...

Page 5: ...bration See section 4 5 on pages 14 15 for temperature calibration Preparing the Meter for Calibration The pH conductivity temperature probe included with this meter is designed for use with this mete...

Page 6: ...ate the pH standard buffer solution See figure NOTE If using a pH buffer other than pH 7 press the or keys to scroll up or down until the secondary display value is the same as your pH buffer value pH...

Page 7: ...containers 2 If necessary press the MODE key to select Conductivity Mode The S or mS indicator will appear on the right side of the display 3 Rinse your probe with deionized water then rinse the probe...

Page 8: ...and the secondary display shows you what the temperature value was initially See figure 4 5 ATC CAL Temp C B A B pH ATC CAL pH pH A 4 Compare the primary display reading to a NIST traceable thermomete...

Page 9: ...electrode has dehydrated soak it for 30 minutes in OAKTON electrode storage solution or a 2M 4M KCl solution 2 Switch on the meter The MEAS annun ciator appears on the top center of the LCD The ATC i...

Page 10: ...a Clean probe See Probe Care Maintenance p 18 18 7 Probe Care and Maintenance Under typical operating conditions the probe will need to be replaced every 6 to 12 months In extreme applications the pr...

Page 11: ...Error Messages LCD Indicates Cause Solution Display Err Unrecognized input Wrong input in Release key Select annunciator from keypad selected mode valid operations depending on mode CAL Err Calibratio...

Page 12: ...D 00653 04 Electrode storage solution 1 pint bottle Keeps electrode bulb moist for faster more accurate readings Shpg wt 1 1 lbs 510 gms WD 00653 06 Electrode cleaning solution 1 pint bottle Removes b...

Reviews: