Temperature Calibration

The built-in temperature sensor included
in the probe is factory calibrated. Calibrate
your sensor only if you suspect tempera-
ture errors that may have occurred over a
long period of time or if you have a
replacement probe.

1. 

Turn the meter on. Press MODE to
select pH Measurement mode.

2. 

Press the CAL/MEAS key to enter pH
calibration mode. 

The CAL indicator

will appear above the primary display.

See figure 

3. 

While in pH calibration mode, press 
the MODE key to enter temperature
calibration mode. 

The primary display

shows the temperature reading with 
zero offset and the secondary display
shows you what the temperature value
was initially.

See figure

4.5

ATC

CAL

Temp

C

B

A

B

pH

ATC

CAL

pH

pH

A

4. 

Compare the primary display reading
to a NIST-traceable thermometer or
another thermometer known to be
accurate. 

5. 

Press the

or

keys

to adjust 

the primary display reading to agree
with your temperature standard.  

See figure

6. 

Press the ENTER key

to confirm tem-

perature calibration. The CON indicator
will flash to confirm calibration.

See figure 

Notes

To exit from Temperature Calibration mode
without confirming calibration, 

DO NOT

press ENTER in step 6. Press CAL/MEAS
instead.

Temperature calibration is restricted to
±5°C from the factory default value 
displayed during calibration

(shown in 

the secondary display).

15

ATC

CAL

Temp

C

C

ATC

CAL

Temp

CON

C

D

D

C

Summary of Contents for 35630-00

Page 1: ...will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Au...

Page 2: ...ting 19 9 Error Messages 20 10 Specifications 21 11 Accessories 22 23 12 Warranty 24 13 Return of Items 24 2 3 pH Conductivity C Meter HOLD MODE ON OFF CAL MEAS C MEAS pH ATC pH Con 10 Series ENTER 1...

Page 3: ...nding graphic indicator on the LCD ON OFF Powers and shuts off the meter HOLD Freezes the measured reading To activate press HOLD while in measurement mode To release press HOLD again MODE Selects the...

Page 4: ...s a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes or electrodes For a replacement probe see the Accessories section page 22 NOTE Keep connector dry and c...

Page 5: ...bration See section 4 5 on pages 14 15 for temperature calibration Preparing the Meter for Calibration The pH conductivity temperature probe included with this meter is designed for use with this mete...

Page 6: ...ate the pH standard buffer solution See figure NOTE If using a pH buffer other than pH 7 press the or keys to scroll up or down until the secondary display value is the same as your pH buffer value pH...

Page 7: ...containers 2 If necessary press the MODE key to select Conductivity Mode The S or mS indicator will appear on the right side of the display 3 Rinse your probe with deionized water then rinse the probe...

Page 8: ...and the secondary display shows you what the temperature value was initially See figure 4 5 ATC CAL Temp C B A B pH ATC CAL pH pH A 4 Compare the primary display reading to a NIST traceable thermomete...

Page 9: ...electrode has dehydrated soak it for 30 minutes in OAKTON electrode storage solution or a 2M 4M KCl solution 2 Switch on the meter The MEAS annun ciator appears on the top center of the LCD The ATC i...

Page 10: ...a Clean probe See Probe Care Maintenance p 18 18 7 Probe Care and Maintenance Under typical operating conditions the probe will need to be replaced every 6 to 12 months In extreme applications the pr...

Page 11: ...Error Messages LCD Indicates Cause Solution Display Err Unrecognized input Wrong input in Release key Select annunciator from keypad selected mode valid operations depending on mode CAL Err Calibratio...

Page 12: ...D 00653 04 Electrode storage solution 1 pint bottle Keeps electrode bulb moist for faster more accurate readings Shpg wt 1 1 lbs 510 gms WD 00653 06 Electrode cleaning solution 1 pint bottle Removes b...

Reviews: