54
How to Respond to Alarms
Alarm Name,
Priority and
Type
Alarm Description
Solution
No Breath
Alarm
(Low Priority)
(Physiological)
When in PULSE mode, if a breath is not
detected for 45 seconds the screen will
turn amber, sound one beep and the text
“No Breath” will flash in the system status
window.
If no breath is detected for an additional
15 seconds, the unit will switch to the
previous CONTINUOUS mode setting
selected.
Ensure cannula tubing
does not exceed 7 feet
(2.1m).Ensure user
breathing through nose
and not mouth. Note: No
Breath Detected is used
when in Pulse Flow Mode
ONLY.
Low Battery
Alert
(Low Priority)
(Technical)
When calculated battery run time
reaches 6 minutes the screen will turn
amber, sound one beep per 30 seconds
and the text “Low Battery” will flash in the
system status window.
Pushing POWER button once silences the
beeping.
Plug into an external
power source.
OR
Replace depleted
battery (batteries) with a
charged battery.
Low Battery
Alarm
(High Priority)
(Technical)
When battery power is depleted, the
pump shuts off, the screen will flash red at
2Hz, sound a 10 beep sequence
repeated after 2.5 seconds of silence and
the text “Low Battery” will flash in the
system status window
The unit will continue to sound until; the
power button is pressed to silence the
audio, the power button is held to power
off the unit, or 15 seconds has expired.
Plug into an external
power source. Replace
depleted battery
(batteries) with a
charged battery.
Low Purity Alert
(Low Priority)
(Technical)
When oxygen levels drop below 85% the
screen will turn amber, sound one beep
per
30 seconds and the text “Low Purity”
will flash in the system status window.
Pushing POWER button once silences the
beeping.
Clean or replace air inlet
filter. Contact your
authorized dealer.
Low Purity
Alarm
(High Priority)
(Technical)
When oxygen levels drop below 72% the
screen will flash red at 2Hz, sound a 10
beep sequence repeated after 2.5
seconds of silence and the text “Low
Purity” will flash in the system status
window
The unit will continue to sound until; the
power button is pressed to silence the
audio, the power button is held to power
off the unit, or 15 seconds has expired.
Repair or replace tubing
or humidifier. Clean or
replace air inlet filter.
Contact your authorized
dealer.
No Flow Alarm
When the flow of oxygen is stopped for 45
Check cannula
connection. Repair or