53
Alarm Name,
Priority and
Type
Alarm Description
How to Simulate
Alarm
(High Priority)
(Technical)
the unit internally) the screen will flash red
at 2Hz, sound a 10 beep sequence
repeated after 2.5 seconds of silence and
the text “High External Power” will flash in
the system status window
The unit will continue to sound until; the
power button is pressed to silence the
audio, the power button is held to power
off the unit, or 15 seconds has expired.
20amps. Run unit on 2.0
continuous. Starting at
14 volts steadily increase
input voltage until unit
alarms.
No External
Power Alarm
(High Priority)
(Technical)
If power is removed while the unit is running
the unit will sound a 10 beep sequence
repeated after 4 seconds of silence.
Since the power is removed, the display is
blank and no backlight color is available.
Run unit on 2.0
continuous using AC
power supply. Remove
all batteries. Remove
AC power supply.
DO NOT IGNORE ALARMS