Using the instrument

Functions

The instrument is operated by three push buttons:

To switch on the instrument and to screen

the main menus and the sub menus.

To enter and activate the menu selected 

by the Select button.

To exit the current menu or function.

The menu

Four main menus are available for selection:

Main menu/Calibrate instructs the instrument to 

measure the White Calibration Device.

Main menu/Enter test contains sub menus correspon-

ding to each of the NycoCard

®

tests that are available.

Main menu/Options contains two sub menus:

Enter language: To change the language on the display.

New white: To measure a new value of white when a

message “Error Recalibrate” appears on the display.

The white value will change with ageing, even if pen

tips and calibration devices are replaced.

Main menu/Off followed by “ENTER” will turn off the

instrument. Pressing QUIT and ENTER simultaneous-

ly will also turn off the instrument, regardless of the

position in the menu.

Operating procedure

The following three steps must be performed before

the result of a NycoCard

®

test is obtained:

1 . Adjusting

Make sure that the pen is placed in 

the penholder and turn the Reader on.

The instrument automatically performs

“Adjusting”. Allow the Adjusting proce-

dure to complete without touching the pen. A single

beep will indicate that the Adjusting is completed 

correctly. A double beep will indicate errors in the

Adjusting followed by the message “Adjust error”.

(See troubleshooting).

2 . White calibration

Press ENTER to activate the menu and

follow the instructions in the display.

When the message “Place pen…” is

displayed, lift the pen out of the holder

and place it on the White Calibration Device. Hold the

pen steady until a single beep indicates that the cali-

bration is finished.

3 . Measuring a NycoCard

®

test

The Reader pad (7) should always be

used during measurements. The reque-

sted NycoCard

®

test is chosen from Main

menu / Enter test and the instrument is

ready for the measurement. When the pen tip is placed

into the test hole and the pen sleeve pushed down, the

measurement will start. After a few seconds the test

result will be displayed as a concentration value.

If the instrument is placed on stand-by for about 10

minutes, it will automatically switch-off. When the

instrument is turned on again, the last measured value

will automatically be displayed.

Maintenance

The Reader pen has two replaceable accessories, the

pen tip (14) and the pen ring (13). These two parts will

be in contact with the assay devices that are exposed

to diluted blood/urine. The pen tip must be replaced if

it is dirty, to reduce interference during measurements.

Be careful not to touch the optics of the Reader pen.

It is recommended that the pen tip and white calibra-

tion device are inspected regularly. If the calibration

device is damaged or dirty, it should be changed.

Replacing the pen tip: to remove the pen tip, care-

fully snap the tip off with your fingers. Replace the

new pen tip.

Replacing the pen ring: to remove the pen ring,

hold the case shield with one hand and carefully

wriggle the ring off with the other hand. Replace

the new pen ring. Make sure it fits correctly.

Replacing the calibration device: remove the device

by simultaneously pressing the small handle down

and the opposite side upwards. Place the new cali-

bration device.

Replacing the batteries:

Step 1. Use a screwdriver to open the battery lid (15).

Step 2. Remove the old batteries. It is recommen-

ded to use a small screwdriver when 

removing the battery clips (17).

Step 3. Replace the batteries with three new 

batteries.

Be sure that the batteries are inserted in

the right direction.

Step 4. Replace the battery lid. New batteries

need to be charged before using the 

instrument.

6

Enter

Select

ON

Quit

Summary of Contents for Reader II

Page 1: ...light I is then inversely proportional to T and the microprocessor converts the time to a concentration value which will then be displayed When the pen is placed over the sample and the pen sleeve is pushed down the light dark detector regis ters that the environment has changed from light to dark and the measurement is automatically started The sleeve will ensure that the pen is held stable in a ...

Page 2: ... Measuring a NycoCard test The Reader pad 7 should always be used during measurements The reque sted NycoCard test is chosen from Main menu Enter test and the instrument is ready for the measurement When the pen tip is placed into the test hole and the pen sleeve pushed down the measurement will start After a few seconds the test result will be displayed as a concentration value If the instrument ...

Page 3: ...atteries has short 3 Replace all batteries circuited and is not recharging Display screen is blank or 1 Batteries discharged see other 1 Recharge the batteries Check unreadable causes under Battery low polarity and that the batteries are of the correct type 2 Sub optimal viewing conditions 2 Increase light level or change light Light too dim position Display fades out during Adjusting 1 Batteries ...

Page 4: ...ecommended range is 18 25 C Relative humidity 0 90 Avoid direct sunlight or too dim light 1 See special instructions for each NycoCard test Batteries Rechargeable NiMH batteries only size AA 1 2 V Recharge 8 18 hours Battery charger Mascot type 8311 0901 or equivalent Output 9VDC 2 5 W Plug 3630 coaxial type Polarity pos in centre Length 12 0 mm Inner diameter 2 1 mm Outer diameter 5 5 mm Ordering...

Reviews: