background image

Introduction to NycoCard

®

READER II

The NycoCard

®

READER II is a small battery powered

instrument, designed to measure all NycoCard

®

products. The coloured responses of the NycoCard

®

products are measured by spectral reflectance in three

parts of the visible spectrum.

Description of NycoCard

®

READER II

The NycoCard

®

READER II kit consists of:

NycoCard

®

READER II

1 unit

Battery charger

1 unit

Batteries (rechargeable)

3 units

Calibration devices

10 units

Pen tips

10 units

Pen rings

10 units

Battery clips

3 units

Pad

1 unit

Instruction manual

1 unit

Warranty card

1 unit

The NycoCard

®

READER II consists of two units: the

instrument box (1), which is the operational and calcu-

lating unit, and the Reader pen (5), which is the mea-

surement unit. The main components of the instrument

box are the microprocessor, the LCD display (3) and

the push buttons (2) for operation. The key elements of

the Reader pen are the Light Emitting Diodes (LED’s)

(11), a photodiode (12), a light/dark detector (10) and

the functional design. A pen cable (4) connects these

two units together. The penholder (8) is located to the

right of the display on the instrument box. The pen is

placed in the holder to protect it when not in use. In

front of this holder, there is a lid (6), and on the rear

side of the lid a white calibration device (9) is placed.

Basic theory

The Reader pen contains three pairs of LED’s: red,

green and blue. Reflected light from the test sample 

is measured, relative to a white sample, by a light-

sensitive photodiode circuit. When a specific test from

the NycoCard

®

READER II test menu is chosen, the

required LED’s illuminate the sample to be measured.

The sample will reflect some of the light and give rise

to a constant current in the photodiode, which will 

discharge a capacitor linearly from the voltage V

1

to 

a lower voltage V

2

. When V

2

is reached, a signal is

given to the microprocessor. The microprocessor 

measures the time (T) it takes to change the voltage

from V

1

to V

2

. The intensity of reflected light (I) is then

inversely proportional to T, and the microprocessor

converts the time to a concentration value, which will

then be displayed.

When the pen is placed over the sample and the pen

sleeve is pushed down, the light/dark detector regis-

ters that the environment has changed from light to

dark and the measurement is automatically started.

The sleeve will ensure that the pen is held stable in 

a vertical position, and that external light is excluded

from the sample.

Adjusting

This procedure compensates for the contribution of

reflected light from the inside of the pen tip (14) and

slight electrical current leakage in the electronic circuit.

Adjusting is performed when the pen is placed in the

penholder, i.e. in a dark chamber without any sample

present. It is important to adjust the instrument each

time the instrument is turned on and when the pen tip

is replaced.

White calibration

White calibration adjusts for any differences in 

measuring conditions and variation of pen tips. The

white calibration is performed on a special calibration

device, which is stored under the lid, protected from

light and dust. The NycoCard

®

READER II should 

be calibrated prior to each series of measurements

(max 20 measurements per series) and prior to each

single, sporadic measurement.

Installation and environment

Check that all components listed in the section

Description of NycoCard

®

READER II are included in

the package and that no components are damaged.

Fill in the Instrument Warranty Card and send it back

to your local distributor.

Make sure that the instrument is acclimatised to room

temperature before use. Recommended ambient 

temperature is 18 – 25°C and humidity must not be

condensing. Ensure that the instrument is protected

from dirt and direct sunlight.

Your NycoCard

®

READER II is supplied with 3 rechar-

geable NiMH batteries (1.2 V) (16). Before using the

new instrument the batteries should be charged for 18

hours. The connection plug of the battery charger (18)

is inserted into the outlet (19) located at the rear side

of the instrument box. The Reader pen has a modular

plug (20). The NycoCard

®

READER II is equipped

with RS 232 port for PC-connection (21).

5

GB

A1

A2

A3

A4

Summary of Contents for Reader II

Page 1: ...light I is then inversely proportional to T and the microprocessor converts the time to a concentration value which will then be displayed When the pen is placed over the sample and the pen sleeve is pushed down the light dark detector regis ters that the environment has changed from light to dark and the measurement is automatically started The sleeve will ensure that the pen is held stable in a ...

Page 2: ... Measuring a NycoCard test The Reader pad 7 should always be used during measurements The reque sted NycoCard test is chosen from Main menu Enter test and the instrument is ready for the measurement When the pen tip is placed into the test hole and the pen sleeve pushed down the measurement will start After a few seconds the test result will be displayed as a concentration value If the instrument ...

Page 3: ...atteries has short 3 Replace all batteries circuited and is not recharging Display screen is blank or 1 Batteries discharged see other 1 Recharge the batteries Check unreadable causes under Battery low polarity and that the batteries are of the correct type 2 Sub optimal viewing conditions 2 Increase light level or change light Light too dim position Display fades out during Adjusting 1 Batteries ...

Page 4: ...ecommended range is 18 25 C Relative humidity 0 90 Avoid direct sunlight or too dim light 1 See special instructions for each NycoCard test Batteries Rechargeable NiMH batteries only size AA 1 2 V Recharge 8 18 hours Battery charger Mascot type 8311 0901 or equivalent Output 9VDC 2 5 W Plug 3630 coaxial type Polarity pos in centre Length 12 0 mm Inner diameter 2 1 mm Outer diameter 5 5 mm Ordering...

Reviews: