Nvis901

pH Meter Calibration using pH buffer solution

Nvis 901 pH meter having three point calibration for correcting pH sensor error. For this three buffer
solution of  

4pH, 7pH and 9.18pH  

is required. The pH buffer solution can be prepared form pH

buffer tabled or by chemical method. However pH buffer tablet are recommended for good accuracy.
Follow the procedure given below for calibration. 

1. Clean the pH sensor by distill water and dip pH sensor in distill water for two minutes.

2. Clean the PT100 sensor by distill water. And dip in pH buffer solution with pH sensor.

3. Now prepare pH buffer solution of 4pH, 7pH and 9.18pH in three different clean beaker.(please see
the procedure given on buffer tablet container to how to prepare pH buffer solution).

4. Now on the instrument and dip pH sensor in 4pH buffer solution and then press enter button two 
times.

”Dip in 4 pH”

 message appear on LCD (Fig 6).

Fig 6

5. Wait for 1 minute and then press 

down 

button. Instrument start taking reading. After few second 

Press  Up

” message appear on LCD(Fig 7). Now press up key.

Fig 7

6. Now LCD display “

Dip in 7pH

”. Dip pH sensor in distill water for few second (without removing

form instrument). Then dip it in to 7pH buffer solution for  

1 minute  

and then press  

down

  button.

Instrument take reading and then message appear “

Dip in 9.18pH”

.

7. Dip pH electrode in distill water  for few second (without removing sensor form instrument). Then
dip in to 9.18pH buffer for one minute. Press down button, “

Press Up

” message appear on display. 

8. Now press up button. The “

Cal Active

” message come on display second line(Fig 8). To activate

calibration press 

Escape 

button and for deactivate calibration press down button, “

Cal Deactivate

message come on display and then press  

Escape

. If calibration deactivate option is selected then

calibration process is not effective.

Nvis Technologies Pvt. Ltd.

11

Summary of Contents for Nvis901

Page 1: ...1 Ver 1 1 Designed Manufactured by 141 B Electronic Complex Pardesipura Indore 452 010 India Tel 91 731 4211500 Telefax 91 731 4202959 Toll free 1800 103 5050 E mail info nvistech comWebsite www nvist...

Page 2: ...Nvis901 Nvis Technologies Pvt Ltd 2...

Page 3: ...Nvis901 Content 1 Introduction 4 2 Features 5 2 Technical Specifications 6 3 Display and Inputs 7 4 Operating Procedure 8 5 PC commutation 13 6 Accessories 20 Nvis Technologies Pvt Ltd 3...

Page 4: ...PC interface 3 point calibration Inbuilt electronic buffer for calibration Auto temperature compensation Manual temperature compensation It provide resolution of 0 01pH and 0 1 C for temperature NV90...

Page 5: ...eatures Micro controller based design Fast response time Resolution of 0 01 pH and 0 1 C for temperature 3 point calibration In built electronic buffers PC software for storing data Nvis Technologies...

Page 6: ...ion 0 01pH Temperature Resolution 0 1 C Electronic Buffer 4 pH 7pH 9 18pH Calibration 3 point PC Interface RS 232 USB Power Supply 230 V 10 50 Hz Automatic Temperature Compensation ATC 0 80 C Manual T...

Page 7: ...nter in menu 4 Escape For Escape from Menu Connectors 1 pH pH Sensor input 2 Temperature Temperature Sensor input Rear Panel Connectors 1 Green Connector 4 pH Electronic Buffer 2 Red Connector 7 pH El...

Page 8: ...Nvis901 5 RS 232 Connector PC Interface 6 USB Connector PC Interface Nvis Technologies Pvt Ltd 8...

Page 9: ...e compensation is enable Fig 2 4 For displaying the electrode potential press UP key for two second and electrode potential appear on the display Fig 3 5 The electrode potential is display till the UP...

Page 10: ...key for decreasing the temperature 3 To out form temperature adjust press ESCAPE key 4 By this MTC is enabled and MTC appear on display in place of ATC Fig 6 Fig 5 5 For disable the MTC press Down ke...

Page 11: ...hen press enter button two times Dip in 4 pH message appear on LCD Fig 6 Fig 6 5 Wait for 1 minute and then press down button Instrument start taking reading After few second Press Up message appear o...

Page 12: ...Always use fresh buffer solution for calibration 3 If given sequence is not followed or buffer are of wrong value then Check the sol message come display 4 pH electrode response time is nearly about...

Page 13: ...ne with pH sensor and buffer solution Electronic calibration not work in case of ph sensor error Electronic calibration procedure The calibration procedure using buffer solution and electronic buffer...

Page 14: ...This software also show duration on which reading taken and plot pH vs time and temperature vs time graph Fig 9 Operating Procedure 1 Install the software on PC 2 Connect Nvis 901 USB probe with PC A...

Page 15: ...ow show duration on which instrument start taking reading Fig 12 Fig 12 5 To start the sampling set the sampling time text box Fig 12 MM is for minute and SS is for second Fig 13 6 Press start button...

Page 16: ...een display the duration of data collection 10 Use Stop button for stopping the data collection 11 Reset button is for resetting the counter and ph and temperature graph 11 Import button display the s...

Page 17: ...odes which fail to respond to the above treatment can be further activated by dipping them to 2 hydrofluoric acid for 5 10 seconds and immediately washing with distilled water This drastic treatment s...

Page 18: ...mperature for long time 4 During MTC setting and calibration process instrument stop sending the data to pc 5 When MTC is active the instrument display the MTC set point and send the MTC set point to...

Page 19: ...mperature Error Table Values in gray are temperature and pH in which there is no error in pH from temperature pH vs mV Table This table show the different value of mV form electrode vs pH Nvis Technol...

Page 20: ...Accessories 1 pH sensor 1No 2 Temperature Sensor 1No 3 Stand 1No 4 Software CD 1No 5 USB Cable 1No 6 RS232 Cable 1No 7 Main Cord 1No 8 Cable for Electronic buffer Calibration 1No Nvis Technologies Pvt...

Page 21: ...municated to us immediately giving full details of the complaints and defects noticed specifically mentioning the type and sr no of the instrument date of purchase etc 6 The repair work will be carrie...

Reviews: