7
SECTION 3
INSTALLATION PROCEDURE
3.1 LIFTING AND TRANSPORT
Because of the heavy weight of the incubator, all lifting and transport must be carried out using
proper handling equipment. The incubator must be supported from underneath and never turned
over.
3.2 UNPACKING
Remove the packing cardboard box and the second nylon packing around the incubator. The
below written are provided with the equipment, please check them;
User
”s manual
1 piece of warranty certification
2 piece of shelves
4 piece of shelf carriers
1 piece of power cable
Check that no damage has occurred during transport
.
3.3 POSITIONING
Lift the incubator underneath and carry it carefully to its place.
Balance the incubator on four pedestals. If necessary, provide stable standing by adjusting the
pedestal heights.
Place the shelf carriers and then the shelves.
Check the followings,
The proposed site is suitable for the user,
The operator can follow up the even he deals with something else.
The incubator does not occupy the utilisation space of others or does not damage them.
Leave at least 20 cm. free space between the equipment and wall.
ATTENTION !!!
Please pay special attention to the followings,
Indoor use only
Temperature from 5°C to 40°C
Maximum relative humidity of 80% for temperature up to 22°C,
Maximum altitude: 2000 m.
The maximum performance is obtained between 15°C and 25°C.
At most 70% of the surface area of the shelves should be used in order to obtain a uniform
temperature distribution.
Summary of Contents for EN 300
Page 14: ...13 SECTION 8 ELECTRICAL CIRCUIT DIAGRAM 8 1 EN 300 ELECTRICAL CIRCUIT DIAGRAM...
Page 15: ...14 8 2 EN 400 ELECTRICAL CIRCUIT DIAGRAM...
Page 16: ...15 8 3 EN 500 ELECTRICAL CIRCUIT DIAGRAM...
Page 17: ...16 8 4 EN 400P ELECTRICAL CIRCUIT DIAGRAM...
Page 18: ...17 8 5 EN 500P ELECTRICAL CIRCUIT DIAGRAM...
Page 19: ...18 SECTION 9 WARNING LABEL...
Page 20: ...19 GROUNDED PLUG EN 300 FUSES 2x1A EN 400 EN 400P FUSES 2x 1 6A EN 500 EN 500P FUSES 2x2A...