Kit L
Operating temperature range
0°C to +55°C (+32°F to +131°F)
Storage temperature range
-20°C to +70°C (-4°F to +158°F)
Vibration (50 Hz ... 2,000 Hz)
≤200 ms
-2
(20 g)
Shock (11 ms)
≤400 ms
-2
(40 g)
Humidity
≤93% relative humidity (no condensation forming)
6.6
Ambient Conditions
6.7 Cable
Scanning Head Cable (connection cable from connector PCB to D-Sub connector)
• Provide a strain relief for the cable near the connector PCB�
• The encoder cables and connecting cables must be laid away from sources of interference (e�g�
mains cables, fuses, motors, magnetic valves or power supplies). Normally a distance of ≥100 mm
will be sufficient.
• Install the cable in a way that it cannot be damaged by the rotating hub� Please pay attention to the
permissible bending radius (see also point 6�2!
Extension Cable (connection cable from D-Sub connector to controller)
• Use preferably original extension cables made by NUMERIK JENA� This guarantees an optimum
compatibility with the encoders and ensures a maximum protection against electromagnetic
interferences�
• Please consult the technical support of NUMERIK JENA before using selfmanufactured extension
cables�
• Do not install any other signal lines in the encoder cable!
Chart 5
18/58
KIT
L
-Series