31
the breast milk container 12
attached to
the breast shield in an upright position.
Caution!
To ensure the required results,
never allow the milk level to rise beyond
the lower rim of the lip valve
11
.
4. To start pumping, touch either of the
buttons
22
,
23
, or
24
.
Pumping will start with the fastest suction
rhythm and the lowest suction strength
(default setting)
.
Setting the suction strength
5. Use the buttons
23
and
24
to
adjust the suction strength to one of
4 levels.
Touch-and-hold will cause the suction
strength to adjust step by step up to the
maximum or minimum level respectively.
The selected suction strength is
displayed with
31
(First Choice
+
) or
38
(Nature Sense).
You can adjust the suction strength any
time during the pumping session.
Setting the suction rhythm
6. Touch the
22
button once to adjust
the suction rhythm to the next level.
Touch-and-hold will cause the suction
rhythm to adjust step by step until you
release the button.
The selected suction rhythm is displayed
with
27
(First Choice
+
) or
32
(Nature
Sense).
You can adjust the suction rhythm any
time during the pumping session.
Pause mode
7. Touch the Pause button
25
to place
the pump unit into pause mode. To
continue pumping, touch the Pause
button
25
again.
Memory function
The memory function can be used to save
the sequence of a pumping session from the
moment it is started.
8. To save a pumping session, you must
first change to pause mode by touching
the Pause button
25
.
9. Then touch the Memory button
21
for at least 2 seconds.
While the session is being saved, the
symbol
28
(First Choice
+
) or
33
(Nature
Sense) will flash.
Once the session has been saved, the
pump will return to start-up mode.
10. To start the saved session, touch the
Memory button
21
while the pump unit
is in start-up mode after being switched on.
The symbol
28
(First Choice
+
) or
33
(Nature
Sense) is lit.
Nature Sense model:
in addition, the
screen background will change to red.
11. You can pause and resume the pumping
session at any time by touching the Pause
button
25
, even while running a saved
session.
12. To exit the Memory function and return to
the default settings for the remainder of the
pumping session, touch either of the
buttons
22
,
23
, or
24
while
the pump unit is in pause mode.
This will return your pumping session to
the default settings.
Troubleshooting
13. If you notice a reduction in the suction
power over time, clean the filter
6
(see
“5.3 Special cleaning of tube and filter“ on
page 29), or swap the clogged filter for the
replacement filter included in the delivery
package.
14. If milk has entered the tube, immediately
switch off the pump unit using the master
power switch
26
, and clean the tube
and the filter
6
(see “5.3 Special cleaning of
tube and filter“ on page 29).