29
• Incorrect treatment can damage the
parts.
• Before using again, allow the sterilised
parts to
cool and dry completely
before
reassembling.
• Only use water and steam for
sterilisation:
1. Use a NUK Steam Steriliser, following the
operating instructions of the device.
Alternatively,
2. sterilise the parts in boiling water for 5
minutes.
Ensure there is sufficient water in the pan
to avoid damage.
Alternatively,
3. use a NUK Micro Express Plus
Microwave Steam Steriliser for
sterilisation in your microwave oven.
• Only use the NUK Micro Express Plus
Microwave Steam Steriliser for
microwave sterilisation. Follow the
operating instructions of the device.
5.3 Special cleaning of tube
and filter
, switch the
device off immediately.
2. Dismantle the device as shown in figure
,
proceeding in reverse from 7 to 1.
3. Remove the filter
(see figure
4. Clean the following parts with lukewarm
water and a little washing-up liquid and/
or NUK Rinsing Agent:
– rinse filter
;
– flush tube
with tube connector plug
5. Reassemble according to figure
C
or
B
in
reverse order and then use the blow-dry
function (see “5.4 Tube blow-dry
function“ on page 29) of the device to dry
the tube
and the tube connector plug
Note:
Since the filter
would be a virtually
impermeable barrier while wet or damp, it
would inhibit the blow-dry function and the
pump suction. Insert the dry replacement
filter instead. Leave the freshly cleaned and
rinsed damp filter to air-dry for at least
4 hours.
5.4 Tube blow-dry function
1. Assemble the device as shown in figure
B
.
Do not insert a filter that is still damp, but
use a dry replacement filter instead.
2. To start the blow-dry function press the
master power switch
and then
simultaneously touch the buttons
and
.
The blow-dry function runs for 5 minutes.
3. Check that the tube is completely dry (no
droplets visible inside), before starting
the next pumping session.
If the tube is not yet completely dry, start
the blow-dry function again.
You can pause the blow-dry function by
touching the Pause button
and
continue it later by touching the Pause
button
again.
4. To stop the blow-dry function, switch the
pump unit off by pressing the master
power switch
.
6 Assembling the pump
1. Assemble the device as shown in figure
B
.
Note:
Make sure the silicone cushion
fits
snugly in the breast shield
and the lip
is properly mounted, as the pump
can otherwise draw in air, which will prevent it
from working properly.