Operating temperature:
+10.0~40.0 °C (50.0~104.0 °F), 15%~85% RH
Storage temperature:
-20.0~+50.0 °C (-4.0~122.0 °F)
relative humidity
≦
85 %
Transport temperature:
<70.0 °C (158.0 °F)
relative humidity
≦
95 %
Accuracy
:
Forehead mode:
±0.2 °C (0.4 °F) within 35~42 °C (95.~107.6 °F),
±0.3 °C (0.5 °F)
for other range
Surface temperature mode:
±0.3 °C (0.5 °F)
in range
22.0~42.2 °C (71.6~108.0 °F), otherwise ±4 %
(however at least ±2.0 °C / 4.0 °F)
Air pressure:
800~1013 hPa
Protection class:
BF
Degree of protection:
IP22
Battery:
1 CR2032 lithium battery
Battery life:
approx. 3,000 measurements (~1 year)
Comply with:
ASTM E1965-98,
EN ISO 80601-2-56,
IEC/EN60601-1-2(EMC),
IEC/EN60601-1(Safety) standards,
ISO10993, RoHS.
Clinical Bias:
-1.4~ -1.7 °C
Limits of Agreement:
0.98
Repeatability:
0.20°C
Expected service life:
3 years from initial use
Note
The NUK Flash Thermometer converts forehead temperature into its "oral
equivalent" (relative to the results of the clinical trial).