BEFORE USE

Low battery capacity is indicated with a flashing red light. We recommend that you 

replace the new batteries if the low battery indicator turns red.

Battery Replacement

Make sure the oximeter is switched off when replacing the batteries.

The oximeter is powered by 4 AA alkaline batteries. Replace new batteries by the 

following steps. 

STEP1. 

Loosen the screw at the back of the oximeter and remove the battery cover.

STEP2. 

Remove the old batteries from battery compartment.

STEP3. 

Insert four new AA alkaline batteries. Correctly align the polarities (+ and -) 

with battery indication marks on the oximeter.

STEP4. 

Replace the battery cover and tighten the screw.  

CAUTION! 

1.Use only 1.5 AA new batteries with this device. Replace the batteries as soon 

as possible after a low battery indicator appears. 

2.If the oximeter does not work after installed the batteries. Reinstall the batteries.

STEP1

STEP4

STEP2

STEP3

Summary of Contents for PO8201

Page 1: ...Model PO8201...

Page 2: ......

Page 3: ...Ver 1 0 May 2010 311 8201100 003 Owner s Manual Model PO8201...

Page 4: ...rnings and notes throughout this manual are very important Please read this entire manual carefully before using the nu beca Handheld Pulse Oximeter The information in the manual has been carefully ch...

Page 5: ...al Description 08 Principle of Measurement 08 Meter Appearance and Key Function 09 Content of the System 11 BEFORE USE 12 Battery Replacement 12 Connecting the Probe 13 Power On Off 13 Set up the Oxim...

Page 6: ...toring sites have trauma disability or other medical status that make inaccurate results operators should consult doctors before use Use probe manufactured by manufacturer Using other manufacturers pr...

Page 7: ...cators illuminate and that an audible indicator sounds during the startup initialization sequence if any indicator is not lit or the audible indicator does not sound do not use the oximeter Please con...

Page 8: ...e range of the peak wavelengths and maximum optical output power of the light emitted by the oximeter probe and the reading result can be especially useful to clinicians The oximeter with reusable SpO...

Page 9: ...Meter Appearance and Key Function Reading Value Display Area 1 2 3 4 9 10 6 7 5 8 Pulse Amplitude 8 Low Battery Indicator SpO2 indicator LED Screen Set Button Advance Button On Off Button SpO2 Pulse...

Page 10: ...p settings and keep the oximeter silence when it is alarming 4 Advance Button Is used to review status or adjust alarm volume 5 On Off Button Is used to turn on or turn off the oximeter by pressing On...

Page 11: ...ulse Oximeter with Probe B Owner s Manual x 1 C AA Alkaline Batteries x 4 D Warranty Card Confirm that the items listed are packed with the nu beca Handheld Pulse Oximeter If any item on this list is...

Page 12: ...sen the screw at the back of the oximeter and remove the battery cover STEP2 Remove the old batteries from battery compartment STEP3 Insert four new AA alkaline batteries Correctly align the polaritie...

Page 13: ...obe to the top of the oximeter as shown below Make sure that the probe is rmly plugged into the oximeter Power On Off Turn on the oximeter by pressing and releasing the button Turn off the oximeter by...

Page 14: ...ase follow below steps Start with the oximeter is off STEP 1 Press Button and button at the same time to enter setting mode The rst screen shows SpO2 upper limit settings The factory default is off ST...

Page 15: ...tton again and the screen will turn back OFF setting STEP 4 Press button to change settings You can press button to increase the number from 85 to 94 STEP 3 When the desired setting is displayed on th...

Page 16: ...number from 105 BPM to 250 BPM When the setting value goes to the upper limit press button again and the screen will turn back OFF setting STEP 5 When the desired setting is displayed on the screen p...

Page 17: ...increase the number from 30 BPM to 100 BPM When the setting value goes to the upper limit press button again and the screen will turn back OFF setting STEP 7 When the desired setting is displayed on t...

Page 18: ...is brighter than L2 setting press button to turn off the oximeter For PO 8201B STEP 11 Press to change settings Press button to turn on Bluetooth STEP 9 When the desired setting is displayed on the sc...

Page 19: ...e Rate Low Off Off 30 100 5 BPM Brightness L1 L1 and L2 N A Power On Self Test When the oximeter is turned on for normal operation the unit will check if the oximeter is ready for use If probe does no...

Page 20: ...e oximeter is ready to use CAUTION 1 Consult health professional before you start to use the oximeter 2 The oximeter probe might not work on cold extremities due to reduced circulation Warm or rub the...

Page 21: ...ate on LED STEP 4 The readings will be updated based on the signal received with each pulse SpO2 Symbol SpO2 The SpO2 symbol shows the current oxygen saturation in percentage Pulse Rate Symbol The pul...

Page 22: ...alarm settings have not been recalled or adjusted in Setup mode the default alarm settings remain in effect Refer to the table below for detailed information about alarm conditions activation criteri...

Page 23: ...ts SpO2 probes can be cleaned and disinfected with same solutions Let the probe dry before using it again The rubber inside of the SpO2 probe belongs to medical rubber which has no toxin and no harmfu...

Page 24: ...or connects to patient Excessive motion at the probe site may be prohibiting the oximeter from acquiring a consistent pulse signal Eliminate or reduce the cause of the motion artifact or reposition th...

Page 25: ...imeter from the EMI environment Displays and indicators are off but a continuous audible alarm is sounding Problem in operation Turn the oximeter off and replace the batteries If the problem still per...

Page 26: ...d representative in the European Community c i n o r t c e l e d n a l a c i r t c e l e r o f n o i t c e l l o c e t a r a p e s s e t a c i d n I equipment WEEE CE Marking indicating conformance to...

Page 27: ...icate a stable reading is taken Probe fall off detection indication Battery status indication Keys On Off key One button Functional key Two buttons Alarm Urgent alarm SpO2 value is higher lower than p...

Page 28: ...122 F for the device 5 C to 40 C 41 F to 104 F for the probe Storage Temperature 20 C to 55 C 4 F to 131 F Operating Humidity Below 95 non condensing Classi cation Degree of Protection Type BF applie...

Reviews: