nubeca FT7122 Instruction Manual Download Page 3

3/11 

 

TECHNICAL SPECIFICATIONS 

Measurement time: 

5 seconds 

Measurement location: 

The area between the end of eyebrow and the hairline 

Measurement range: 

35.0°C ~ 43.0°C   

Accuracy: 

±0.2°C – 35.0°C ~ 35.4°C 
±0.1°C – 35.5°C ~ 42.0°C 
±0.2°C – 42.1°C ~ 43.0°C 

Display unit: 

0.1°C 

Storage (transport) 
condition: 

-25°C ~ 55°C (-13°F ~ 131°F ) / 30% - 85% RH / 
700 – 1060hpa

 

Operation condition: 

15°C ~ 35°C 

(59°F ~ 95°F ) / 10% – 95% RH    / 700 – 

1060hpa

 

Battery:     

One Lithium CR2032 battery (3V) 

Battery life:     

1,000 measurements approximately (under normal use) 

Weight: 

Approx. 25 grams including battery 

Dimensions: 

93.65 mm (L) x 34.75 mm (W) X 24.85 mm(T) 

Memory: 

Last reading display when turn on 

Applied part: 

  type BF (sensor)

 

Degree of protection 
from liquid and solid 
particle: 

IP 22 

The thermometer is not waterproof. 

Note 1: The thermometer is an adjusted mode clinical thermometer 
Note 2: This thermometer converts the forehead temperature to display its “oral 
equivalent”. 

 

°C / °F SWITCH 

To switch between 

 and 

, press and keep pressing the on/off button for 5 seconds 

when power off, then the LCD screen will show the default unit. Press the button again 
to switch to the opposite unit. 

 

MEASURING GUIDELINES 

1.  Measurement location is the area between the end of eyebrow and the hairline. 
2.  Do not place the thermometer on scarred tissue compromised by skin disorder, 

open sores or abrasions and patients in trauma. 

3.  Temperature readings are affected by many factors including exercise and drinking 

prior to taking a reading. 

4.  Consult your physician when using drug therapies that may raise the local skin 

temperature in the forehead; this may lead to incorrect reading. 

Summary of Contents for FT7122

Page 1: ...13 Applying Standards 7 Section 14 Symbol Explanation 7 Section 14 EMC Table 8 Read the instruction manual thoroughly and carefully before using the unit What is normal body temperature Normal body t...

Page 2: ...ce is not intended to be a substitute for consultation with your physician 3 Disregarding the information given in this instruction manual is considered to be abnormal use 4 The device is for indoor u...

Page 3: ...article IP 22 The thermometer is not waterproof Note 1 The thermometer is an adjusted mode clinical thermometer Note 2 This thermometer converts the forehead temperature to display its oral equivalent...

Page 4: ...d it against the measurement location to ensure good contact with the skin is maintained Correct placing and good contact with the skin ensure reliable readouts Make sure the thermometer and the patie...

Page 5: ...Drawings Instructions 1 Gently press the top of the battery cover so as to release the battery cover 2 Slide back the battery cover 3 Take the battery out When placing a new battery please make sure t...

Page 6: ...is included in the LOT The LOT has 5 numbers The first number means the year 0 9 The second number means the month Jan Sep 1 9 Oct A Nov B Dec C The third and fourth numbers mean the date 1 31 And the...

Page 7: ...ANDARDS 1 MDD 93 42 EEC 2 EN 60601 1 2006 AC 2010 3 EN 60601 1 2 2007 4 EN ISO 80601 2 56 2012 5 ASTM E1112 00 Symbol Explanation The CE mark and Notified Body Registration Numbers the requirement of...

Page 8: ...hat supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Guidance and manufacturer s declar...

Page 9: ...ations equipment should be used no closer to any part of the FT7122 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitte...

Page 10: ...broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered I...

Page 11: ...o 800 MHz d 1 2 800 MHz to 2 5 GHz d 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommen...

Reviews: