OM0240
|
Rev 6 December
/2019
Page 22 of 34
NuAire, Inc.
|
2100 Fernbrook Lane
|
Plymouth, MN 55447
|
U.S.A
|
ph: 763.553.1270
|
fx: 763.553.0459
|
tf: 800.328.3352
Postcool
ON
OFF
Pre-cooling Mode
AIR
ROTOR
Temperature display mode
AIR
ROTOR
Enter service menu and enter password : (9876)
LOCKING TEMPO
UNLOCKING TEMPO
IMBALANCE SETTING
4.5 Replacement of Cylinders
4.5.1 Ventilated Model
Open the lid.
Switch the centrifuge off & disconnect it.
Remove the c-clip & pin located at the top of the cylinder using c-clip pliers.
Gently swivel the lid backward.
Unscrew the 3 screws located under the bowl seal.
Disassemble the bowl by pulling toward the top, and then remove it.
Hold the lid in the open position
Remove the screw located at the bottom of the cylinder
Remove the defective cylinder holding the lid open.
Install the new cylinder, with the spiral spring down.
Insert the screw at the bottom of the cylinder.
Reposition the bowl & its 3 screws.
Place the pin & the snap rings on top of the cylinder, in the lid.
Use the new bolts during the reassembly of the bowl.
4.5.2 Refrigerated Model
Access the rear of the centrifuge and clear access to the cylinder, from the right side.
Remove the c-clip & pin located at the top of the cylinder using c-clip pliers.
Gently swivel the lid backward.
Remove the screw located at the bottom of the cylinder.
Remove the faulty cylinder.
Install the new cylinder.
Insert the screw at the bottom of the cylinder.
Place the pin & the c-clip on top of the cylinder, in the lid.