OM0284
Page 45 of 72
Rev 2 July/2021
The cabinet is considered to be certifiable if the following airflow measurements are present:
a. Downflow average: 60 LFPM
5 LFPM (.30 m/s
.025 m/s).
b. Inflow average:
105 LFPM
5 LFPM (.53 m/s
.025 m/s) using the direct inflow measurement
method or constricted 3 inch (76mm) high access opening measurement method.
Note:
It is recommended to perform the airflow calibration process at least annually to optimize both the airflow
control and Intelli
Flow™ monitor systems.
Summary of Contents for LabGard NU-543-300E
Page 7: ...OM0284 Page 7 of 72 Rev 2 July 2021...
Page 9: ...OM0284 Page 9 of 72 Rev 2 July 2021...
Page 10: ...OM0284 Page 10 of 72 Rev 2 July 2021...
Page 11: ...OM0284 Page 11 of 72 Rev 2 July 2021...
Page 12: ...OM0284 Page 12 of 72 Rev 2 July 2021...
Page 16: ...OM0284 Page 16 of 72 Rev 2 July 2021...
Page 21: ...OM0284 Page 21 of 72 Rev 2 July 2021 Figure 1...
Page 22: ...OM0284 Page 22 of 72 Rev 2 July 2021...
Page 35: ...OM0284 Page 35 of 72 Rev 2 July 2021...
Page 67: ...OM0284 Page 67 of 72 Rev 2 July 2021...
Page 68: ...OM0284 Page 68 of 72 Rev 2 July 2021...
Page 69: ...OM0284 Page 69 of 72 Rev 2 July 2021...
Page 70: ...OM0284 Page 70 of 72 Rev 2 July 2021...