10.0 Troubleshooting
Should there be any malfunctions in the system, please attempt to answer the following questions before
notifying maintenance or calling Nuaire. Please do not dismantle the freezer yourself.
Fault
Troubleshooting
Freezer does not
start up
Is the power supply normal? Has the main protector been turned on yet?
Is the voltage supply too low?
Is there any voltage input from the outside?
Poor
refrigeration
effect
Is the ambient temperature too high?
Are the inner doors and outer doors closed properly? (Has any ice or frost damaged the
seals between the door and the frame?)
Is the condenser filter clogged?
Is the temperature setting correct?
Is the freezer being kept away from direct exposure to sunlight?
Is the freezer near any heat source?
Is the porthole plug installed in the porthole with proper insulation materials?
Has the freezer been loaded with too many non-frozen samples within the last few
hours?
The unit is noisy
Is the unit set on a firm and level floor?
Is the exterior of the unit touching any objects?
Is the freezer unit leveled with the leveling legs?