12
13
(4) Changing Lamp (Varios 560 LUX)
1) Refer to 3. (5) Handpiece Connection. Disassemble the handpiece from the Handpiece Code. Remove the cover.
Use a precision screw driver and push lamp out.
2) Align the Lamp Pins of a new lamp with the holes, and push the lamp into.
* Optional Lamp : VA Bulb Set (A Set consist of 3 pieces) Order No. Y900107
Fig. 24
Cover
Lamp
Socket
View from A
Lamp Pin
Lamp
A
3) Align the Mark on the cover with the same on the Handpiece
Cord. Push the cover to the cord unit it clicks. Assemble the
handpiece and the Handpiece Cord. (Refer. 3. (5) Handpiece
Connection)
Fig. 25
Mark
Cover
Handpiece Cord
Mark
CAUTION
· Make sure that the bulb is sufficient cooled, wheen it is
replaced.
· Do not touch glass part of new lamp.
· Care needs to be execised when mounting a new lamp
to avoid the O-ring off from the groove or being twisted.
(5) Maintenance of the Control Unit etc
If dirt sticks on the Control Unit, Foot Control, or Handpiece Cord, remove the Power Cord first. Then, soak a clean
cloth in water and squeeze cloth tightly to drain water. Use the cloth to wipe away dirt. After that, wipe off with an
alcohol soaked clean cloth.
When trouble is found, check the following items again before asking repairs. If none of these are applicable or the trouble
is not remedied even after action has been taken, a failure of this product is suspected. Contact your dealer.
9. Troubleshooting
No / Poor
vibration.
The tip is bent
or broken.
The tip walks
out.
Noise from the
handpiece.
The handpiece
is overheating.
The Power Lamp
does not light, even
if the Power Switch
is ON.
The tip does not
generate vibration,
in spite of
depressing the Foot
Control.
The Power Cord or the Jack is
disconnected.
The internal Fuse has burned out.
The tip is not tightened firmly.
Worn Tip.
Output has not been correctly
adjusted for the Tip.
The Foot Control is disconnected.
Failure of vibrator in the handpiece.
Failure of the Foot Control.
Output has not been properly
adjusted for the Tip.
The Tip is not tightened firmly.
Output has not been properly
adjusted for the Tip.
The Tip is not tightened firmly.
Failure of vibration in the
handpiece or the Control Unit.
Output has not been properly
adjusted for the Tip.
The Tip is not tightened firmly.
Failure of vibration in the
handpiece or the Control Unit.
Correctly insert the Power Cord or the Jack.
Contact your dealer.
Tighten the Tip until the Tip Wrench rotates
freely with clicks.
Replace the Tip.
Adjust the Power on the Tip Case. Do not
exceed.
Connect the Foot Control correctly.
Contact your dealer.
Contact your dealer.
Adjust the power on the Tip Case. Do not
exceed.
Tighten the tip until the Tip Wrench rotates
freely with clicks.
Adjust the power on the Tip Case. Do not
exceed.
Tighten the Tip until the Tip Wrench rotates
freely with clicks.
Contact your dealer.
Adjust the power on the Tip Case. Do not
exceed.
Tighten the tip until the Tip Wrench rotates
freely with clicks.
Contact your dealer.
Problem
Probable Cause
Cause
Solution